A compact and hand-held infection-screening system for use in rapid medical inspection at airport quarantine stations: system design and preliminary validation
- 26 February 2015
- journal article
- Published by Taylor & Francis in Journal of Medical Engineering & Technology
- Vol. 39 (3) , 185-190
Abstract
To conduct mass screening and thereby reduce the spread of infection, a compact (13.5 cm × 8.5 cm × 2.5 cm), highly-mobile and hand-held infection-screening system was developed for rapid medical inspection in mass gathering places such as airports. The system is capable of non-contact vital-sign monitoring using two integrated sensors: a 24-GHz microwave radar for measuring heart and respiration rates and a thermopile array for capturing facial temperature. Subsequently, the system detects infected individuals using a linear discriminant function (LDA) from the derived vital-signs data. The system was tested on 10 subjects under two conditions (resting as normal and exercising as pseudo-infected, i.e. a 10-min bicycle ergometer at 100 W exercise); the normal and pseudo-infected conditions were classified successfully via LDA for all subjects (p < 0.01; classification error rate < 5%). The proposed non-contact system can be applied for preventing secondary exposure of medical doctors at the outbreak of highly pathogenic infectious diseases such as the Ebola virus.Keywords
This publication has 23 references indexed in Scilit:
- Emerging Infectious Diseases: Threats to Human Health and Global StabilityPLOS Pathogens, 2013
- Screening for Influenza A(H1N1)pdm09, Auckland International Airport, New ZealandEmerging Infectious Diseases, 2012
- Fever screening during the influenza (H1N1-2009) pandemic at Narita International Airport, JapanBMC Infectious Diseases, 2011
- Entry screening to delay local transmission of 2009 pandemic influenza A (H1N1)BMC Infectious Diseases, 2010
- Development of a non-contact screening system for rapid medical inspection at a quarantine depot using a laser Doppler blood-flow meter, microwave radar and infrared thermographyJournal of Medical Engineering & Technology, 2009
- Antioxidant supplementation in sepsis and systemic inflammatory response syndromeCritical Care Medicine, 2007
- Motion Artifact Reduction in Photoplethysmography Using Independent Component AnalysisIEEE Transactions on Biomedical Engineering, 2006
- Entry screening for severe acute respiratory syndrome (SARS) or influenza: policy evaluationBMJ, 2005
- Limitations of Forehead Infrared Body Temperature Detection for Fever Screening for Severe Acute Respiratory SyndromeInfection Control & Hospital Epidemiology, 2004
- The challenge of emerging and re-emerging infectious diseasesNature, 2004