Analytic Performance of Immunoassays for Drugs of Abuse Below Established Cutoff Values

Authors:

Abstract
Background: The analytic performance and accuracy of drug detection below Substance Abuse and Mental Health Services Administration (SAMHSA) cutoffs is not well known. In some patient populations, clinically significant concentrations of abused drugs in urine may not be detected when current SAMHSA cutoffs are used. Our objectives were to define the precision profiles of three immunoassay systems for drugs of abuse and to evaluate the accuracy of testing at concentrations at which the CV was <20%.