Red cell damage after pumping by two infusion control devices (Arcomed VP 7000 and IVAC 572)
Authors:
- 31 July 2007
- journal article
- research article
- Published by Wiley in Transfusion Medicine
- Vol. 17 (4) , 290-295
Abstract
summary The aim of this study was to assess the performance in terms of red cell damage of two peristaltic volumetric infusion pumps – the Alaris IVAC 572 (San Diego, CA, USA) and Arcomed Volumed VP7000 (Regensdorf, Switzerland).Various infusion pumps are available to transfuse blood at a predetermined rate. It is recommended that each machine should be individually assessed.This experiment used six units of single‐donor‐transfusable packed red cells and ran each unit through both pumps. This was carried out at 9, 28 and 35 days post‐donation at rates from 2 to 150 mL h−1. Post‐pumping samples from these experiments, and a pre‐pumping sample in each case, were analysed for levels of potassium and free haemoglobin (Hb). They were also examined microscopically for evidence of cell damage.Potassium levels showed no significant change with pumping on any occasion, but rose significantly as the samples aged. Free Hb showed some variation, but the only consistent finding was a similar rise in value with increasing pack age. Microscopic examination revealed no cell damage under any condition.Both pumps performed to an acceptable level and appear safe to be used for red cell transfusion.Keywords
This publication has 12 references indexed in Scilit:
- Changes in red blood cell integrity related to infusion pumps: A comparison of three different pump mechanismsPediatric Critical Care Medicine, 2003
- The effect of external pressure, catheter gauge, andstorage time on hemolysis in RBC transfusionTransfusion, 2001
- The administration of blood and blood components and the management of transfused patientsTransfusion Medicine, 1999
- Haemolysis following rapid experimental red blood cell transfusionActa Anaesthesiologica Scandinavica, 1998
- Red cell viability with infusion systemsTransfusion, 1994
- Evaluation of a volumetric intravenous fluid infusion pump for transfusion of blood components containing red cellsTransfusion, 1986
- Effects of intravenous delivery systems on infused red blood cellsAmerican Journal of Health-System Pharmacy, 1984
- Does transfusion using a syringe infusion pump and small‐gauge needle cause hemolysis?Transfusion, 1981
- Administration of Microaggregate Filtered Blood Using a Manual Infusion PumpTransfusion, 1980
- Red Blood Cell Damage by Shear StressBiophysical Journal, 1972