Robust Fat Suppression for High‐Resolution DWI at 5 T Using Slice‐Selection Gradient Modulation and Chemical Shift Encoding
- 22 December 2025
- journal article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 95 (5) , 2594-2610
Abstract
Purpose: Chemical shift encoding (Dixon) can simultaneously remove multiple fat peaks in multi‐shot EPI diffusion‐weighted imaging (ms‐ EPI DWI ) at 3 T with low sensitivity to B 0 inhomogeneity. However, this method is not applicable at 5 T , since increased fat off‐resonance frequencies cause slice position mismatches between different fat peaks. In this work, we propose a two‐step strategy combining slice‐selection gradient modulation ( SSGM ) and Dixon to enhance fat suppression for ms‐ EPI DWI at 5 T. Methods: In the first step, SSGM adjusts the amplitudes of excitation and refocusing slice‐selection gradients according to the off‐resonance frequencies of methyl/methylene fat peaks, so that these fat slices are excited but not refocused. In the second step, the olefinic fat peak is chemical shift encoded and separated from the diffusion‐weighted water images through a joint water/fat separation algorithm with structured low‐rank regularization. The two‐step strategy was evaluated in the leg, head‐and‐neck, and prostate. Results: In vivo experiments demonstrated that the Dixon‐only methods cannot simultaneously suppress all fat peaks at 5 T , while this problem was addressed by combining SSGM and Dixon. SSGM showed superior suppression for methyl/methylene fat compared to SPAIR . The following Dixon further removed olefinic fat untouched by SPAIR . Qualitative analysis showed improved overall image quality for all anatomies. Prostate experiments showed that the proposed method is also applicable in reduced FOV acquisitions, high‐resolution (1.6‐mm isotropic) and high b value imaging (2800 s/mm 2 ). Conclusion: The proposed two‐step strategy improved fat suppression in ms‐ EPI DWI at 5 T , which can potentially enhance whole‐body disease screening and diagnosis.Keywords
Funding Information
- National Key Research and Development Program of China (2022‐YFC‐2405303)
This publication has 73 references indexed in Scilit:
- Quantification of the Proton Density Fat Fraction and Iron Content: A Comparative Study Between 3.0 T and 5.0 T MRIJournal of Magnetic Resonance Imaging, 2025
- Evaluating the Diagnostic Performance of MR Cytometry Imaging in Differentiating Benign and Malignant Breast TumorsJournal of Magnetic Resonance Imaging, 2025
- Diffusion MRI of the breast: Current status and future directionsJournal of Magnetic Resonance Imaging, 2019
- Motion artifact reduction of diffusion‐weighted MRI of the liver: Use of velocity‐compensated diffusion gradients combined with tetrahedral gradientsJournal of Magnetic Resonance Imaging, 2012
- Relaxation times of skeletal muscle metabolites at 7TJournal of Magnetic Resonance Imaging, 2009
- Improved fat suppression using multipeak reconstruction for IDEAL chemical shift fat‐water separation: Application with fast spin echo imagingJournal of Magnetic Resonance Imaging, 2009
- Evaluation of optimized inversion‐recovery fat‐suppression techniques for T2‐weighted abdominal MR imagingJournal of Magnetic Resonance Imaging, 2008
- Multicoil Dixon chemical species separation with an iterative least‐squares estimation methodMagnetic Resonance in Medicine, 2003
- Why fat is bright in rare and fast spin‐echo imagingJournal of Magnetic Resonance Imaging, 1992
- Three‐point dixon technique for true water/fat decomposition with B0 inhomogeneity correctionMagnetic Resonance in Medicine, 1991