Wireless powering and data communication for neural implantable electrodes
- 1 June 2013
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- Vol. 47, 148-153
Abstract
A high resolution RFID reader, collecting neural data from implanted electrodes is proposed. The system is powering the implant using a class E power amplifier (PA) and in between receives the data by an asynchronous demodulation. The technology used is a standard 65 nm CMOS TSMC. Simulations reveal an average power consumption of the overall system of 19 mW with a supply of 1.2 V at 300 MHz of carrier frequency and a BER of 0.3%.Keywords
This publication has 5 references indexed in Scilit:
- Implantable biomedical devices: Wireless powering and communicationIEEE Communications Magazine, 2012
- A 0.013 ${\hbox {mm}}^{2}$, 5 $\mu\hbox{W}$, DC-Coupled Neural Signal Acquisition IC With 0.5 V SupplyIEEE Journal of Solid-State Circuits, 2011
- 0.8 μW 12-bit SAR ADC sensors interface for RFID applicationsMicroelectronics Journal, 2010
- A 5 $\mu$W/Channel Spectral Analysis IC for Chronic Bidirectional Brain–Machine InterfacesIEEE Journal of Solid-State Circuits, 2008
- A 65nm CMOS 30dBm class-E RF power amplifier with 60% power added efficiency2008