A Case of a C-Stem Fracture at the Head-Neck Junction and a Review of the Literature
Authors:
Open Access
- 9 December 2012
- journal article
- review article
- Published by Hindawi Limited in Case Reports in Orthopedics
- Vol. 2012, 1-4
Abstract
We report the first case of a fracture of the standard C-stem in combination with a large metal-on-metal articulation. This occurred at the head-neck junction. Analysis of the fractured stem showed evidence of fatigue failure with possible corrosion. The use of large femoral heads with neck adaptors and narrow tapers should be used with caution, especially in heavy, active patients.Keywords
This publication has 22 references indexed in Scilit:
- Corrosion at the neck-stem junction as a cause of metal ion release and pseudotumour formationThe Journal of Bone and Joint Surgery. British volume, 2012
- Corrosion-Induced Fracture of a Double-Modular Hip ProsthesisJournal of Bone and Joint Surgery, 2010
- Early Results of Conversion of a Failed Femoral Component in Hip Resurfacing ArthroplastyJournal of Bone and Joint Surgery, 2007
- Friction of total hip replacements with different bearings and loading conditionsJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2006
- In vitro comparison of frictional torque and torsional resistance of aged conventional gamma-in-nitrogen sterilized polyethylene versus aged highly crosslinked polyethylene articulating against head sizes larger than 32 mmActa Orthopaedica, 2006
- Early fatigue failures of cemented, forged, cobalt-chromium femoral stems at the neck–shoulder junctionThe Journal of Arthroplasty, 2001
- In vivo corrosion of modular hip prosthesis components in mixed and similar metal combinations. The effect of crevice, stress, motion, and alloy couplingJournal of Biomedical Materials Research, 1993
- Fracture of the Femoral Prosthesis in Total Hip Replacement According to CharnleyActa Orthopaedica, 1977
- Fracture of Femoral Prostheses in Total Hip ReplacementClinical Orthopaedics and Related Research, 1975
- Factors in the Mechanical Failure of the Femoral Component in Total Hip Prosthesis:Report of Six Fatigue Fractures of the Femoral Stern and Results of Experimental Loading TestsActa Orthopaedica, 1974