Variation of the impact duration during the in vitro insertion of acetabular cup implants
Authors:
- Adrien MichelAMAdrien Michel
- Charles-Henri Flouzat LachanietteCLCharles-Henri Flouzat Lachaniette
- Alexandre PoignardAPAlexandre Poignard
- Philippe HernigouPHPhilippe Hernigou
- Jérome AllainJAJérome Allain
- Guillaume HaïatGHGuillaume Haïat
- 6 June 2013
- journal article
- research article
- Published by IOP Publishing in Medical Engineering & Physics
- Vol. 35 (11) , 1558-1563
Abstract
The acetabular cup (AC) is an implant impacted into a bone cavity and used for hip prosthesis surgery. Initial stability of the AC is an important factor for long term surgical success. The aim of this study is to determine the variations of the impact duration during AC implant insertion. Twenty-two bone samples taken from bovine femurs were prepared ex vivo for the insertion of an acetabular cup implant, following the surgical procedure used in the clinic. For each bone sample, ten impacts were applied using reproducible mass falls (3.5 kg) in order to insert the AC implant. Each impact duration was recorded using a wide bandwidth force sensor. For all bone samples, the impact duration was shown to first decrease as a function of the impact number, then reaching a stationary value equal in average to 4.2±0.7 ms after an average number of 4.1±1.7 impacts. The impact duration may be related to variations of the bone-implant interface contact rigidity because of an increase the amount of bone tissue in contact with the AC implant. Measurements of impact duration have a good potentiality for clinical application to assist the surgeon during the insertion of the AC implant, providing valuable information on the bone-implant interface contact properties.Keywords
This publication has 43 references indexed in Scilit:
- Influence of Healing Time on the Ultrasonic Response of the Bone-Implant InterfaceUltrasound in Medicine & Biology, 2012
- The influence of acetabular cup material on pelvis cortex surface strains, measured using digital image correlationJournal of Biomechanics, 2012
- Assessment of MRI issues at 3-Tesla for metallic surgical implants: findings applied to 61 additional skin closure staples and vessel ligation clipsJournal of Cardiovascular Magnetic Resonance, 2012
- MAGNETIC RESONANCE IMAGING SUSCEPTIBILITY ARTIFACTS DUE TO METALLIC FOREIGN BODIESVeterinary Radiology & Ultrasound, 2011
- Ultrasonic Evaluation of Dental Implant Biomechanical Stability: An In Vitro StudyUltrasound in Medicine & Biology, 2011
- In vivo evaluation of a vibration analysis technique for the per-operative monitoring of the fixation of hip prosthesesJournal of Orthopaedic Surgery and Research, 2009
- Effects of screw eccentricity on the initial stability of the acetabular cupInternational Orthopaedics, 2006
- The Norwegian Arthroplasty Register: 11 years and 73,000 arthroplastiesActa Orthopaedica, 2000
- Failed innovation in total hip replacement: Diagnosis and proposals for a cureActa Orthopaedica, 1993
- LOWER RISK OF THROMBOEMBOLIC DISEASE AFTER TOTAL HIP REPLACEMENT WITH NON-CEMENTED THAN WITH CEMENTED PROSTHESESThe Lancet, 1986