Surface Acoustic Wave-Based Microfluidic Coagulation Device for Monitoring Anticoagulant Therapy
- 19 December 2018
- journal article
- review article
- Published by Georg Thieme Verlag KG in Seminars in Thrombosis and Hemostasis
- Vol. 45 (03) , 253-258
Abstract
An universal coagulation test that reliably detects prolonged coagulation times in patients, regardless of which anticoagulant is administered, is not yet available. The authors developed a novel, miniaturized device utilizing surface acoustic waves (SAW) to detect clotting, and used it to develop a novel universal microfluidic coagulation test. Results from this assay were compared with results from standard coagulation assays to detect classical anticoagulants (unfractionated heparin, argatroban) and direct oral anticoagulants (dabigatran, rivaroxaban). The SAW-clotting time (SAW-CT) of this novel device was prolonged in a dose-dependent manner for heparin, argatroban, dabigatran, and rivaroxaban, comparable to standard assays. The authors confirmed the clinical utility of this device in a small patient population admitted to a stroke unit. Preliminary clinical data prove the suitability of the SAW-CT in patients receiving warfarin, rivaroxaban, or dabigatran. The device could be particularly useful as a point-of-care test to assess whole blood coagulation (e.g., in stroke units or in other emergency settings).Keywords
This publication has 8 references indexed in Scilit:
- Microfluidic coagulation assay for monitoring anticoagulant therapy in acute stroke patientsThrombosis and Haemostasis, 2017
- Micro-optical prototyping of a surface acoustic wave-based point-of-care coagulation assay and first application in anticoagulated patientsInt. Journal of Clinical Pharmacology and Therapeutics, 2016
- Emergency admissions for major haemorrhage associated with direct oral anticoagulantsThrombosis Research, 2015
- A novel μ-fluidic whole blood coagulation assay based on Rayleigh surface-acoustic waves as a point-of-care method to detect anticoagulantsBiomicrofluidics, 2013
- Laboratory testing of rivaroxaban in routine clinical practice: When, how, and which assaysAnnals of Medicine, 2013
- Recommendations for the emergency management of complications associated with the new direct oral anticoagulants (DOACs), apixaban, dabigatran and rivaroxabanClinical Research in Cardiology, 2013
- Oral Anticoagulants – A Frequent Challenge for the Emergency Management of Acute Ischemic StrokeCerebrovascular Diseases, 2012
- Shear-induced unfolding triggers adhesion of von Willebrand factor fibersProceedings of the National Academy of Sciences of the United States of America, 2007