Abstract
Since the International Commission on Radiological Protection, in publication ICRP60, has issued new recommendations on radiation protection quantities, there is now increasing interest in commercially available instruments optimized and calibrated for the measurement of neutron ambient dose equivalent H*(10). Therefore within a joint cooperation agreement between the Research Center Karlsruhe and EG&G Berthold, the new neutron-dose-rate meter LB6411 was developed. The detector system with integrated electronics has a 3 He proportional counter tube centered in a moderating sphere. The response between thermal energies and 20 MeV was optimized with the help of extensive MCNP Monte-Carlo calculations. The instrument has an extremely high sensitivity of approximately 3 counts per nSv and can be used both as a portable or as a stationary neutron monitor. Fluence responses have been measured in monoenergetic neutron beams provided by the Physikalisch-Technische Bundesanstalt (PTB) in Braunschweig, Germany. The ambient dose equivalent response of the LB6411 is reported over the whole energy range. As the ratio of new to the old conversion function is a function of neutron energy, a simple recalibration of existing remcounters by a trivial factor is not considered to be sufficient. Therefore within a joint cooperation agreement between the Research Center Karlsruhe and EG&G Berthold, the new neutron dose rate meter LB6411 was developed. There had been two major objectives for the new design. Firstly, it was intended to tune the energy dependent response to ambient dose equivalent H*(10) from thermal energies up to 20 MeV, and secondly an improved neutron detection efficiency in comparison with existing instruments should be achieved.