Abstract
Conventional otoscopes and oto-endoscopes, which are used to examine the tympanic membrane (TM), do not provide tomographic information. Optical coherence tomography (OCT) non-invasively reveals the depth-resolved internal microstructure of the TM with very high spatial resolution. We designed this study to examine the TMs with middle ear diseases using a handheld otoscope employing 860 nm spectral domain (SD)-OCT, combined with video camera and to demonstrate the clinical applicability of this system. A total of 120 patients with otologic symptoms were enrolled. TM images were obtained using the handheld OCT-based otoscope (860 nm central wave length, 15 μm axial resolution, 15 μm lateral resolution, and 7 mm scanning range using relay lens). Both OCT and oto-endoscope images were compared according to the clinical characteristics such as perforation, retraction, and postoperative healing process. The objective grade about the thickness of perforation margins and the accurate information about the extent of TM retraction that was not distinguishable by oto-endoscopic exam could be identified using this system. The postoperative healing process of TMs could be also followed using the OCT device. These analyses from the surgeon-oriented perspective suggest another useful application of the handheld OCT device.
Funding Information
  • Ministry of Trade, Industry & Energy (10047943)
  • U.S. National Institutes of Health (1 R01 EB013723)
  • National Research Foundation of Korea (NRF-2015R1C1A1A02036457)

This publication has 20 references indexed in Scilit: