Abstract
OBJECTIVE. The purposes of this study were to directly compare 3D endoluminal volume rendering and navigational capabilities of three different CT colonography systems and to assess feasibility of 3D evaluation for primary polyp detection.MATERIALS AND METHODS. Closely matched endoluminal images from three CT colonography software systems (Navigator, Vitrea 2, and V3D-Colon) and optical colonoscopy were obtained of eight pathologically proven colorectal polyps. All images were then reviewed by 25 physicians (12 radiologists and 13 gastroenterologists) who were not familiar with the three systems. For each polyp, the images yielded by the three systems were rated according to polyp conspicuity, 3D effect (depth), subjective quality, and likeness to optical colonoscopy. For comparison of endoluminal navigation capability, automated or semiautomated flight from rectum to cecum and cecum to rectum was attempted in 10 cases (20 flights) in which a continuous colonic air column could be identified on the 2D ima...