Evaluation of Indicator-Based pH Measurements for Freshwater over a Wide Range of Buffer Intensities
- 3 July 2008
- journal article
- research article
- Published by American Chemical Society (ACS) in Environmental Science & Technology
- Vol. 42 (16) , 6092-6099
Abstract
Two different sulfonephthalein indicators, cresol red (CR) with a pKa of ∼8.3 and bromothymol blue (BTB) with pKa of ∼7.4, were tested for an analysis of freshwater over a broad range of pH and total alkalinity values. Measurements from an autonomous sensor system using a 1 cm optical path length were compared to those using a 10 cm path length on a benchtop spectrophotometer. The indicator pH perturbation was quantified with a thermodynamic model and nonlinear least-squares analysis. The laboratory study found that the perturbation-corrected pH differed between the 1 cm (large indicator perturbation) and 10 cm (small indicator perturbation) optical path length measurements from −0.017 to +0.15 with a median of +0.0041 pH units for CR and from −0.015 to +0.026 with a median of −0.0008 pH units for BTB. Precision was ±0.0005−0.013 and ±0.0001−0.0027 pH units for the 1 and 10-cm-path-length measurements, respectively. The autonomous sensor was deployed for 14 days in a local creek. Simultaneous glass pH electrode measurements had a large negative and drifting offset (−0.15 to −0.40 pH units) compared to the indicator-based measurements. This study is the first in situ comparison between potentiometric and spectrophotometric pH methods in a freshwater system.Keywords
This publication has 24 references indexed in Scilit:
- Spectrophotometric pH measurements of freshwaterAnalytica Chimica Acta, 2002
- An Evaluation of In-Situ Measurements of Water Temperature, Specific Conductance, and pH in Low Ionic Strength StreamsWater, Air, & Soil Pollution, 1998
- Diel, episodic and seasonal changes in pH and concentrations of inorganic carbon in a productive lakeFreshwater Biology, 1996
- Shipboard flow injection determination of sea water pH with spectrophotometric detectionAnalytica Chimica Acta, 1995
- The internal consistency of CO2 measurements in the equatorial PacificMarine Chemistry, 1993
- Electrode pH error, seasonal epilimnetic pCO2, and the recent acidification of the maine lakesWater, Air, & Soil Pollution, 1990
- High precision multiwavelength pH determinations in seawater using cresol redDeep Sea Research Part A. Oceanographic Research Papers, 1989
- Spectrophotometric determination of seawater pH using phenol redAnalytical Chemistry, 1985
- Performance tests for the measurement of pH with glass electrodes in low ionic strength solutions including natural watersAnalytical Chemistry, 1985
- Determination of hydrogen ion concentration in softwater lakes using carbon dioxide equilibriaGeochimica et Cosmochimica Acta, 1984