A Computer Model of the Pediatric Circulatory System for Testing Pediatric Assist Devices
Authors:
- 1 January 2007
- journal article
- pediatric mechanical-circulatory-support
- Published by Wolters Kluwer Health in ASAIO Journal
- Vol. 53 (1) , 74-81
Abstract
Lumped parameter computer models of the pediatric circulatory systems for 1- and 4-year-olds were developed to predict hemodynamic responses to mechanical circulatory support devices. Model parameters, including resistance, compliance and volume, were adjusted to match hemodynamic pressure and flow waveforms, pressure-volume loops, percent systole, and heart rate of pediatric patients (n = 6) with normal ventricles. Left ventricular failure was modeled by adjusting the time-varying compliance curve of the left heart to produce aortic pressures and cardiac outputs consistent with those observed clinically. Models of pediatric continuous flow (CF) and pulsatile flow (PF) ventricular assist devices (VAD) and intraaortic balloon pump (IABP) were developed and integrated into the heart failure pediatric circulatory system models. Computer simulations were conducted to predict acute hemodynamic responses to PF and CF VAD operating at 50%, 75% and 100% support and 2.5 and 5 ml IABP operating at 1:1 and 1:2 support modes. The computer model of the pediatric circulation matched the human pediatric hemodynamic waveform morphology to within 90% and cardiac function parameters with 95% accuracy. The computer model predicted PF VAD and IABP restore aortic pressure pulsatility and variation in end-systolic and end-diastolic volume, but diminish with increasing CF VAD support.Keywords
This publication has 16 references indexed in Scilit:
- Physiological Control of Blood Pumps Using Intrinsic Pump Parameters: A Computer Simulation StudyArtificial Organs, 2006
- The National Heart, Lung, and Blood Institute Pediatric Circulatory Support ProgramCirculation, 2006
- Left Ventricular and Myocardial Perfusion Responses to Volume Unloading and Afterload Reduction in a Computer SimulationASAIO Journal, 2004
- Control Strategy for Maintaining Physiological Perfusion with Rotary Blood PumpsArtificial Organs, 2003
- Mechanical circulatory support for infants and children with cardiac diseaseThe Annals of Thoracic Surgery, 2002
- Pediatric Circulatory Support SystemsASAIO Journal, 2002
- Modeling and Control of a Brushless DC Axial Flow Ventricular Assist DeviceASAIO Journal, 2002
- Aortic input impedance in infants and childrenJournal of Applied Physiology, 2000
- Estimation of Timing Errors for the Intraaortic Balloon Pump Use in Pediatric PatientsASAIO Journal, 1999
- Evaluation of Ventricular Function in Infants and ChildrenPediatric Clinics of North America, 1971