Comparative Biomechanical Analysis of an Improved Novel Pedicle Screw With Sheath and Bone Cement
Authors:
- Tomoyuki TakigawaTTTomoyuki Takigawa
- Masato TanakaMasato Tanaka
- Hitoshi KonishiHKHitoshi Konishi
- Hisanori IkumaHIHisanori Ikuma
- Haruo MisawaHMHaruo Misawa
- Yoshihisa SugimotoYSYoshihisa Sugimoto
- Kazuo NakanishiKNKazuo Nakanishi
- Koichi KuramotoKKKoichi Kuramoto
- Keiichiro NishidaKNKeiichiro Nishida
- Toshifumi OzakiTOToshifumi Ozaki
- 1 August 2007
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Spinal Disorders & Techniques
- Vol. 20 (6) , 462-467
Abstract
A human cadaveric biomechanical study of fixation strength of an improved novel pedicle screw (NPS) with cement and a conventional screw. To clarify whether the NPS has adequate fixation strength without leakage in vertebrae with low bone quality. The fixation strength of pedicle screws decreases in frail spines of elderly osteoporotic patients. Augmentation of screw fixation with bone cement must be balanced against increased difficulty of screw removal and risk of cement leakage. We developed the NPS consisting of an internal screw and an outer sheath to mitigate the disadvantages of cement augmentation. The T12 and L1 vertebrae obtained from 18 formalin preserved cadavers (11 males and 7 females; mean age, 82.7 y) were used. The mean bone mineral density was 0.39±0.14 g/cm2. The NPS was inserted into one pedicle of each vertebra and the control screw, a Compact CD2 screw, was inserted into the contralateral pedicle. Both screws were 6 mm in diameter and 40 mm in length. Pull-out tests were performed at a crosshead speed of 10 mm/min. Cyclic loading tests were performed with a maximum 250 N load at 2 Hz until 30,000 cycles. Cement leakage did not occur in any of the specimens tested. The mean maximum force at pull-out was 760±344 N for the NPS and 346±172 N for the control screw (P<0.01). Loosening of 50% of the screws was observed after 17,000 cycles of the NPS and after 30 cycles of the control screw. The hazard ratio of loosening was 19.6 (95% confidence interval 19.3-19.9) (P<0.001). The NPS showed a significantly higher mechanical strength than the control screw in both pull-out tests and cyclic loading tests. The NPS showed more than adequate strength without cement leakage.Keywords
This publication has 28 references indexed in Scilit:
- Evaluation of Cyanoacrylate Augmentation of Transpedicular Screw Pullout StrengthJournal of Spinal Disorders & Techniques, 2005
- Efficacy of novel-concept pedicle screw fixation augmented with calcium phosphate cement in the osteoporotic spineJournal of Orthopaedic Science, 2005
- Loads on an internal spinal fixation device during walkingJournal of Biomechanics, 1997
- Effect of formaldehyde fixation on some mechanical properties of bovine boneBiomaterials, 1995
- Stability of Transpedicle Screwing for the Osteoporotic SpineSpine, 1993
- Vertebral fracture assessment using a semiquantitative techniqueJournal of Bone and Mineral Research, 1993
- Pedicle Screw Pullout StrengthSpine, 1993
- Mechanical properties of microcallus in human cancellous boneJournal of Orthopaedic Research, 1992
- Importance of Bone Mineral Density in Instrumented Spine FusionsSpine, 1991
- Loads on the lumbar trunk during level walkingJournal of Orthopaedic Research, 1989