Increased incidence of electrical abnormalities in a pacemaker lead family
- 10 February 2021
- journal article
- research article
- Published by Wiley in Journal of Cardiovascular Electrophysiology
- Vol. 32 (4) , 1111-1121
Abstract
Introduction: Several recent studies have raised concern about noise detections on Tendril pacemaker leads, which may represent insulation failure or algorithm‐driven overreporting of physiologic signals.Methods: We identified all pacemaker leads actively followed at Albany Medical Center, of which 1111 leads (262 Abbott Tendril, 576 Medtronic CapSure, 30 Fineline, 195 Ingevity, 48 Dextrus) in 703 patients were included in this observational study. Electrical abnormalities, comprising low‐impedance measurements n = 50; 19.1%) compared with CapSure (n = 9; 1.6%), Fineline (n = 0), Ingevity (n = 0), and Dextrus (n = 4; 8.3%) leads. The risk of abnormalities was significantly higher in Tendril leads (HR 9.6, 95% CI 5.2–17.6; p < .001). Low impedances were measured on 19 Tendril leads, a significantly higher risk than on other leads (HR 23.8, 95% CI 5.5–102.1; p < .001). Although observation and reprogramming sensitivity were the initial management strategy for 45 and 7 leads, respectively, 18 ultimately required surgical intervention, including 15 Tendrils. No electrical abnormalities were observed in 12 non‐Tendril leads attached to Abbott devices compared with 48 of 252 Tendrils attached to Abbott devices (log‐rank p = .035).Conclusion: Tendril leads demonstrate significantly higher risk of repeated low impedances and noise compared to other manufacturers' models, raising concern that these findings reflect early insulation failure. Increased scrutiny is warranted.Keywords
This publication has 15 references indexed in Scilit:
- Increased Incidence of Noise in the Tendril Pacemaker Lead Detected via Remote MonitoringHeart, Lung and Circulation, 2020
- Detection of high-frequency artifact as a function of pulse generator algorithms and outer-insulation materialHeart Rhythm, 2019
- Long-term performance of a pacing lead family: A single-center experienceHeart Rhythm, 2019
- Pacemaker lead surveillance and failure: Is there a signal in the noise?Heart Rhythm, 2018
- Deaths caused by the failure of Riata and Riata ST implantable cardioverter-defibrillator leadsHeart Rhythm, 2012
- Riata implantable cardioverter-defibrillator lead failure: Analysis of explanted leads with a unique insulation defectHeart Rhythm, 2012
- Causes of ventricular oversensing in implantable cardioverter-defibrillators: Implications for diagnosis of lead fractureHeart Rhythm, 2010
- Time-Dependent Risk of Fidelis Lead FailureThe American Journal of Cardiology, 2010
- Increasing hazard of Sprint Fidelis implantable cardioverter-defibrillator lead failureHeart Rhythm, 2009
- Outcome of the Fidelis implantable cardioverter-defibrillator lead advisory: A report from the Canadian Heart Rhythm Society Device Advisory CommitteeHeart Rhythm, 2008