A Low Voltage Rail-to-Rail OPAMP Design for Biomedical Signal Filtering Applications

Authors:
  • HCHwang-Cherng Chow
    Hwang-Cherng Chow
  • PWPu-Nan Weng
    Pu-Nan Weng

Abstract
In this paper, a 1 volt rail-to-rail input range amplifier has been proposed. In the amplifier design, the current driven bulk (CDB) technique is adopted to reduce threshold voltages of input devices to eliminate the conventional dead zone problem. The performance results of the proposed 1 volt low power amplifier are input rail-to-rail with gain at 71 dB and CMRR at 100 dB. We use the proposed amplifier to implement a low pass filter for biomedical signal applications. Simulation results for an electrocardiograph (ECG) signal are very satisfactory. The total harmonic distortion of the low pass filter is -30 dB. Besides, the programmable cut-off frequency of the bio-signal low pass filter is provided for use.

This publication has 1 reference indexed in Scilit: