Optimization of Precision and Accuracy in X-Ray Fluorescence Analysis of Silicate Rocks
Author:
- 1 July 1985
- journal article
- research article
- Published by SAGE Publications in Applied Spectroscopy
- Vol. 39 (4) , 697-703
Abstract
The precision of XRF measurements on a Rigaku 3064 has been determined for 10 major and 12 trace elements in 10 geochemical reference rocks. A complete procedure for evaluating the sources of imprecision in XRF analyses is presented with new equations and data demonstrating that for many measurements a minute is an adequate count time for obtaining the minimum error (1% RSD) in an analysis. Longer analysis times were found to be limited by sample preparation errors, and for very long analysis times, the precision is predicted to get worse.Keywords
This publication has 16 references indexed in Scilit:
- Energy‐dispersive x‐ray fluorescence analysis of silicate rocks for major and trace elementsX-Ray Spectrometry, 1984
- Comprehensive major and trace element analysis of geological material by x‐ray fluorescence, using low dilution fusionsX-Ray Spectrometry, 1982
- Experimental and Theoretical Comparison of the Precision of Flame Atomic Absorption, Fluorescence, and Emission MeasurementsApplied Spectroscopy, 1981
- Studies in “Standard Samples” for Use In the General Analysis of Silicate Rocks and Minerals*Geostandards Newsletter, 1980
- Optimization of instrumental variables in flame atomic absorption spectrometryAnalytica Chimica Acta, 1979
- XRF determination of trace and major elements using a single-fused discChemical Geology, 1978
- Some disadvantages of Pt95Au5 as a container for molten silicatesMineralogical Magazine, 1977
- Precision of a commercial, wavelength‐dispersive X‐ray fluorescence spectrometer. A case studyX-Ray Spectrometry, 1977
- Precision of an x-ray fluorescence spectrometric determinationAnalytical Chemistry, 1974
- Sensitivity and limit of detection in quantitative spectrometric methodsJournal of Chemical Education, 1974