Poly[hydroxyethyl acrylate-co-poly(ethylene glycol) diacrylate] Monolithic Column for Efficient Hydrophobic Interaction Chromatography of Proteins
Authors:
- 19 October 2009
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 81 (22) , 9416-9424
Abstract
Rigid poly[hydroxyethyl acrylate-co-poly(ethylene glycol) diacrylate] monoliths were synthesized inside 75 μm i.d. capillaries by one-step UV-initiated copolymerization using methanol and ethyl ether as porogens. The optimized monolithic column was evaluated for hydrophobic interaction chromatography (HIC) of standard proteins. Six proteins were separated within 20 min with high resolution using a 20 min elution gradient, resulting in a peak capacity of 54. The effect of gradient rate and initial salt concentration on the retention of proteins were investigated. Mass recovery was found to be greater than 96%, indicating the biocompatibility of this monolith. The monolith was mechanically stable and showed nearly no swelling or shrinking in different polarity solvents. The preparation of this in situ polymerized acrylate monolithic column was highly reproducible. The run-to-run and column-to-column reproducibilities were less than 2.0% relative standard deviation (RSD) on the basis of the retention times of protein standards. The performance of this monolithic column for HIC was comparable or superior to the performance of columns packed with small particles.Keywords
This publication has 34 references indexed in Scilit:
- Novel in situ polymerized coatings for hydrophobic interaction chromatography mediaJournal of Chromatography A, 2007
- Organic polymer monoliths as stationary phases for capillary HPLCJournal of Separation Science, 2004
- Preparation and HPLC applications of rigid macroporous organic polymer monolithsJournal of Separation Science, 2004
- Preparation of hydrophobic interaction chromatographic packings based on monodisperse poly(glycidylmethacrylate-co-ethylenedimethacrylate) beads and their applicationJournal of Chromatography A, 2004
- Hydrophobic interaction chromatography of proteinsJournal of Biotechnology, 2001
- Continuous rods of macroporous polymer as high-performance liquid chromatography separation mediaAnalytical Chemistry, 1992
- Continuous beds for standard and micro high-performance liquid chromatographyJournal of Chromatography A, 1991
- High-performance liquid chromatography on continuous polymer bedsJournal of Chromatography A, 1989
- Some general aspects of hydrophobic interaction chromatographyJournal of Chromatography A, 1973
- Salting-out in Amphiphilic Gels as a New Approach to Hydrophobia AdsorptionNature, 1973