Underestimation of sulfate concentration in PM2.5 using a semi-continuous particle instrument based on ion chromatography
- 23 March 2009
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in Journal of Environmental Monitoring
- Vol. 11 (6) , 1292-1297
Abstract
Recently, a variety of ion-chromatography-based semi-continuous particle instruments such as the Dionex gas particle ion chromatograph (GP-IC), wet-annular denuder/steam-jet aerosol collector (WAD/SJAC), particle-into-liquid sampler with ion chromatograph (PILS-IC), gas and aerosol monitoring system (GAMS) have been introduced for measuring particle chemical components in the atmosphere. It has been reported that sulfate concentrations in PM2.5 measured by these semi-continuous particle instruments correlate well with those measured by other semi-continuous instruments such as the Aerodyne aerosol mass spectrometer (AMS), R&P 8400S and R&P 8400N analyzers, and Thermo model 5020 sulfate particle analyzer and at times exhibit a unity slope. However, the sulfate concentration measured by some of these semi-continuous instruments has been reported to be underestimated by 20–50%, as compared to that in PM2.5 filter samples. In this study, numerous potential causes for underestimation of the sulfate concentrations by the GP-IC were investigated. We found a 30–40% negative artifact arising from an improper calibration procedure inherent within the instrument design. An improved calibration procedure was developed and it substantially reduced the sulfate concentration difference between the GP-IC measurements and PM2.5 filter samples from 36% to 5%.Keywords
This publication has 13 references indexed in Scilit:
- Evaluation and application of a semi-continuous chemical characterization system for water soluble inorganic PM2.5 and associated precursor gasesEGUsphere, 2008
- Water-soluble inorganic ions in airborne particulates from the nano to coarse mode: a case study of aerosol episodes in southern region of TaiwanEnvironmental Geochemistry and Health, 2007
- On the performance of a semi-continuous PM2.5 sulphate and nitrate instrument under high loadings of particulate and sulphur dioxideAtmospheric Environment, 2007
- Measurement of Both Nonvolatile and Semi-Volatile Fractions of Fine Particulate Matter in Fresno, CAAerosol Science and Technology, 2006
- Multi-year urban and rural semi-continuous PM2.5 sulfate and nitrate measurements in New York state: Evaluation and comparison with filter based measurementsAtmospheric Environment, 2006
- A Continuous Analyzer for Soluble Anionic Constituents and Ammonium in Atmospheric Particulate MatterEnvironmental Science & Technology, 2003
- Intercomparison and evaluation of four semi-continuous PM2.5 sulfate instrumentsAtmospheric Environment, 2003
- The performance of a gas and aerosol monitoring system (GAMS) for the determination of acidic water soluble organic and inorganic gases and ammonia as well as related particles from the atmosphereAtmospheric Environment, 2001
- The continuous analysis of nitrate and ammonium in aerosols by the steam jet aerosol collector (SJAC): extension and validation of the methodologyAtmospheric Environment, 2001
- Development of a Sharp-Cut Cyclone for Ambient Aerosol Monitoring ApplicationsAerosol Science and Technology, 2000