Performance characterization of a novel hybrid dosimetry insert for simultaneous spatial, temporal, and motion‐included dosimetry for MR‐linac
Authors:
Open Access
- 13 December 2023
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 51 (4) , 2983-2997
Abstract
Background: Several (online) adaptive radiotherapy procedures are available to maximize healthy tissue sparing in the presence of inter/intrafractional motion during stereotactic body radiotherapy (SBRT) on an MR‐linac. The increased treatment complexity and the motion‐delivery interplay during these treatments require MR‐compatible motion phantoms with time‐resolved dosimeters to validate end‐to‐end workflows. This is not possible with currently available phantoms.Purpose: Here, we demonstrate a new commercial hybrid film‐scintillator cassette, combining high spatial resolution radiochromic film with four time‐resolved plastic scintillator dosimeters (PSDs) in an MRI‐compatible motion phantom.Methods: First, the PSD's performance for consistency, dose linearity, and pulse repetition frequency (PRF) dependence was evaluated using an RW3 solid water slab phantom. We then demonstrated the MRI4D scintillator cassette's suitability for time‐resolved and motion‐included quality assurance for adapt‐to‐shape (ATS), trailing, gating, and multileaf collimator (MLC) tracking adaptations on a 1.5 T MR‐linac. To do this, the cassette was inserted into the Quasar MRI4D phantom, which we used statically or programmed with artificial and patient‐derived motion. Simultaneously with dose measurements, the beam‐gating latency was estimated from the time difference between the target entering/leaving the gating window and the beam‐on/off times derived from the time‐resolved dose measurements.Results: Experiments revealed excellent detector consistency (standard deviation 0.6%), dose linearity (R2 = 1), and only very low PRF dependence (0.4%). The dosimetry cassette demonstrated a near‐perfect agreement during an ATS workflow between the time‐resolved PSD and treatment planning system (TPS) dose (0%–2%). The high spatial resolution film measurements confirmed this with a 1%/1‐mm local gamma pass‐rate of 90%. When trailing patient‐derived prostate motion for a prostate SBRT delivery, the time‐resolved cassette measurements demonstrated how trailing mitigated the motion‐induced dose reductions from 1%–17% to 1%–2% compared to TPS dose. The cassette's simultaneously measured spatial dose distribution highlighted the dosimetric gain of trailing by improving the 3%/3‐mm local gamma pass‐rates from 80% to 97% compared to the static dose. Similarly, the cassette demonstrated the benefit of real‐time adaptations when compensating patient‐derived respiratory motion by showing how the TPS dose was restored from 2%–56% to 0%–12% (gating) and 1%–26% to 1%–7% (MLC tracking) differences. Larger differences are explainable by TPS‐PSD coregistration uncertainty combined with a steep dose gradient outside the PTV. The cassette also demonstrated how the spatial dose distributions were drastically improved by the real‐time adaptations with 1%/1‐mm local gamma pass‐rates that were increased from 8 to 79% (gating) and from 35 to 89% (MLC tracking). The cassette‐determined beam‐gating latency agreed within 12 ms with the ground truth latency measurement. Film and PSD dose agreed well for most cases (differences relative to TPS dose 4%), while film‐PSD coregistration uncertainty caused relative differences of 5%–8%.Conclusions: This study demonstrates the excellent suitability of a new commercial hybrid film‐scintillator cassette for simultaneous spatial, temporal, and motion‐included dosimetry.Keywords
Funding Information
- Nederlandse Organisatie voor Wetenschappelijk Onderzoek (17515)
This publication has 36 references indexed in Scilit:
- Dosimetry in clinical static magnetic fields using plastic scintillation detectorsRadiation Measurements, 2013
- On the nature of the light produced within PMMA optical light guides in scintillation fiber-optic dosimetryPhysics in Medicine & Biology, 2013
- CyberKnife with Tumor Tracking: An Effective Treatment for High-Risk Surgical Patients with Single Peripheral Lung MetastasesFrontiers in Oncology, 2012
- Multichannel film dosimetry with nonuniformity correctionMedical Physics, 2011
- Task Group 142 report: Quality assurance of medical acceleratorsa)Medical Physics, 2009
- Dosimetry for the MRI accelerator: the impact of a magnetic field on the response of a Farmer NE2571 ionization chamberPhysics in Medicine & Biology, 2009
- Plastic scintillation dosimetry and its application to radiotherapyRadiation Measurements, 2006
- A new scintillator detector system for the quality assurance of60Co and high-energy therapy machinesPhysics in Medicine & Biology, 1994
- Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: I. Physical characteristics and theoretical considerationsPhysics in Medicine & Biology, 1992
- Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: II. Properties and measurementsPhysics in Medicine & Biology, 1992