Long-term stability in continuous wave cavity ringdown spectroscopy experiments
- 8 March 2010
- journal article
- Published by Optica Publishing Group in Applied Optics
- Vol. 49 (8) , 1378-1387
Abstract
Allan variance has been used to characterize the slow drift of a near-IR distributed feedback laser-based continuous wave cavity ringdown spectroscopy (CW-CRDS) system. Long-term drift in the cavity loss rate, highly correlated with changes in ambient pressure but not temperature, is observed. With differential measurement of on- and off-peak decay rates, the drift between them largely cancels out, but some residual drift remains if the lasers are detuned more than a few hundred megahertz from each other. A sensitivity to bulk cavity loss () of has been obtained during an optimum integration time of with our CW-CRDS setup, which corresponds to the methane detection limit () in of 0.24 parts in by volume (ppbv) at or 29 parts in by volume (pptv) at pressure. The stability of our system is demonstrated by measuring sub-ppbv methane in at through recording the spectrum of methane lines with our setup.
Keywords
This publication has 12 references indexed in Scilit:
- A broadband absorption spectrometer using light emitting diodes for ultrasensitive, in situ trace gas detectionReview of Scientific Instruments, 2008
- Incoherent broadband cavity enhanced absorption spectroscopy for in situ measurements of NO2 with a blue light emitting diodeApplied Physics B, 2008
- Effects of linear birefringence and polarization-dependent loss of supermirrors in cavity ring-down spectroscopyApplied Optics, 2008
- The use of the Allan deviation for the measurement of the noise and drift performance of microwave radiometersMeasurement Science and Technology, 2007
- Noise in cavity ring-down spectroscopy caused by transverse mode couplingOptics Express, 2007
- High precision measurements of atmospheric nitrous oxide and methane using thermoelectrically cooled mid-infrared quantum cascade lasers and detectorsSpectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy, 2004
- Trace Moisture Detection Using Continuous-Wave Cavity Ring-Down SpectroscopyAnalytical Chemistry, 2003
- The limits of signal averaging in atmospheric trace-gas monitoring by tunable diode-laser absorption spectroscopy (TDLAS)Applied Physics B, 1993
- Cavity ring-down optical spectrometer for absorption measurements using pulsed laser sourcesReview of Scientific Instruments, 1988
- Statistics of atomic frequency standardsProceedings of the IEEE, 1966