Validation of an RP-LC Method for the Determination of Interferon-?2a in Pharmaceutical Formulations

Authors:

Abstract
A reversed phase liquid chromatography (RP-LC) method was validated for the determination of interferon-α2a in pharmaceutical formulations. The RP-LC method was carried out on a Jupiter C4 column (250 mm × 4.6 mm I.D.), maintained at room temperature. The mobile phase A consisted of 0.1% TFA and the mobile phase B was acetonitrile with 0.1% TFA, run at a flow rate of 1 mL/min, and using photodiode array (PDA) detection at 214 nm. The chromatographic separation was obtained with the retention time of 32.6 min, and was linear in the concentration range of 0.5–50 MIU/mL (r 2 = 0.9999). The specificity was proven through degradation studies, which also showed that there was no interference of the excipients. The accuracy was 100.84% with bias lower than 1.87%. The limits of detection and quantitation were 0.19 and 0.5 MIU/mL, respectively. Moreover, method validation demonstrated acceptable results for precision and robustness. The proposed method was applied for the analysis of the interferon-α2a and their related proteins in parenteral dosage forms, contributing to establishing alternatives to improve the quality control assuring the therapeutic efficacy of the biological medicine.

This publication has 19 references indexed in Scilit: