Quantification of lesion size, depth, and uptake using a dual‐head molecular breast imaging system
- 13 March 2008
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 35 (4) , 1365-1376
Abstract
A method to perform quantitative lesion analysis in molecular breast imaging (MBI) was developed using the opposing views from a novel dual‐head dedicated gamma camera. Monte Carlo simulations and phantom models were used to simulate MBI images with known lesion parameters. A relationship between the full widths at 25%, 35%, and of the maximum of intensity profiles through lesions and the true lesion diameter as a function of compressed breast thickness was developed in order to measure lesion diameter. Using knowledge of compressed breast thickness and the attenuation of gamma rays in soft tissue, a method was developed to measure the depth of the lesion to the collimator face. Using the measured lesion diameter and measurements of counts in the lesion and background breast region, relative radiotracer uptake or tumor to background ratio ( ratio) was calculated. Validation of the methods showed that the size, depth, and ratio can be accurately measured for a range of small breast lesions with ratios between 10:1 and 40:1 in breasts with compressed thicknesses between 4 and 10 cm. Future applications of this work include providing information about lesion location in patients for performing a biopsy of site and the development of a threshold for the ratio that can distinguish benign from malignant disease.Keywords
Funding Information
- National Institutes of Health (R21 CA110162)
This publication has 23 references indexed in Scilit:
- Molecular Breast Imaging: Advantages and Limitations of a Scintimammographic Technique in Patients with Small Breast TumorsThe Breast Journal, 2007
- X-ray stereotactic lesion localization in conjunction with dedicated scintimammographyIEEE Transactions on Nuclear Science, 2003
- Phantom study of radiotracer concentration quantification in breast scintigraphyIEEE Transactions on Nuclear Science, 2003
- Small field of view scintimammography gamma camera integrated to a stereotactic core biopsy digital X-ray systemIEEE Transactions on Nuclear Science, 2002
- Evaluation of breast tumor detectability with two dedicated, compact scintillation camerasIEEE Transactions on Nuclear Science, 2002
- 99Tcm-tetrofosmin scintimammography for detecting breast cancer: a comparative study with 99Tcm-MIBINuclear Medicine Communications, 2001
- Optimization of dedicated scintimammography procedure using detector prototypes and compressible phantomsIEEE Transactions on Nuclear Science, 2001
- Multi-PSPMT scintillation cameraIEEE Transactions on Nuclear Science, 1999
- High resolution CsI(Tl)/Si-PIN detector development for breast imagingIEEE Transactions on Nuclear Science, 1998
- Design and assessment of a scintigraphy-guided biplane localization technique for breast tumoursNuclear Medicine Communications, 1996