Diagnosis of feline filariasis assisted by a novel semi-automated microfluidic device in combination with high resolution melting real-time PCR
Authors:
- Achinya PhuakrodAPAchinya Phuakrod
- Witsaroot SripumkhaiWSWitsaroot Sripumkhai
- Wutthinan JeamsaksiriWJWutthinan Jeamsaksiri
- Pattaraluck PattamangPPPattaraluck Pattamang
- Ekachai JuntasaroEJEkachai Juntasaro
- Therdthai ThienthongTTTherdthai Thienthong
- Suporn FoongladdaSFSuporn Foongladda
- Paul J. BrindleyPBPaul J. Brindley
- Sirichit WongkamchaiSWSirichit Wongkamchai
Open Access
- 8 April 2019
- journal article
- research article
- Published by Springer Nature in Parasites & Vectors
- Vol. 12 (1) , 159
Abstract
The diagnosis of filariasis traditionally relies on the detection of circulating microfilariae (mf) using Giemsa-stained thick blood smears. This approach has several limitations. We developed a semi-automated microfluidic device to improve and simplify the detection of filarial nematodes. The efficiency and repeatability of the microfluidic device was evaluated. Human EDTA blood samples were ‘spiked’ with B. malayi mf at high, moderate, and low levels, and subsequently tested 10 times. The device was also used for a field survey of feline filariasis in 383 domesticated cats in an area of Narathiwat Province, Thailand, the endemic area of Brugia malayi infection. In the control blood arbitrarily spiked with mf, the high level, moderate level and low level mf-positive controls yielded coefficient variation (CV) values of 4.44, 4.16 and 4.66%, respectively, at the optimized flow rate of 6 µl/min. During the field survey of feline filariasis in Narathiwat Province, the device detected mf in the blood of 34 of 383 cats (8.9%) whereas mf were detected in 28 (7.3%) cats using the blood smear test. Genomic DNA was extracted from mf trapped in the device after which high-resolution melting (HRM) real-time PCR assay was carried out, which enabled the simultaneous diagnosis of filarial species. Among the 34 mf-positive samples, 12 were identified as B. malayi, 15 as Dirofilaria immitis and 7 as| D. repens. We developed a semi-automated microfluidic device to detect mf of filarial parasites that could be used to diagnose lymphatic filariasis in human populations. This novel device facilitates rapid, higher-throughput detection and identification of infection with filariae in blood samples.Keywords
Funding Information
- Faculty of Medicine Siriraj Hospital MAhidol University (RO 15936006)
- National Science and Technology Development Agency, Ministry of Science and Technology, Thailand (RO15936006, P1551260)
This publication has 27 references indexed in Scilit:
- Integrated Printed Circuit Board Device for Cell Lysis and Nucleic Acid ExtractionAnalytical Chemistry, 2012
- A microfluidic device for whole-animal drug screening using electrophysiological measures in the nematode C. elegansLab on a Chip, 2012
- What is new about animal and human dirofilariosis?Trends in Parasitology, 2009
- Identifying sequence variants in the human mitochondrial genome using high-resolution melt (HRM) profilingHuman Mutation, 2009
- Microfluidic sample preparation: cell lysis and nucleic acid purificationIntegrative Biology, 2009
- Sample preparation: the weak link in microfluidics-based biodetectionBiomedical Microdevices, 2008
- Detection of filarial parasites in domestic cats by PCR-RFLP of ITS1Veterinary Parasitology, 2006
- Selection of surfactants for cell lysis in chemical cytometry to study protein‐DNA interactionsElectrophoresis, 2006
- Discrimination between six species of canine microfilariae by a single polymerase chain reactionVeterinary Parasitology, 2006
- High-Resolution DNA Melting Analysis for Simultaneous Mutation Scanning and Genotyping in SolutionClinical Chemistry, 2005