Transient response of circular foil heat-flux gauges to radiative fluxes
Authors:
- 1 September 1975
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 46 (9) , 1161-1166
Abstract
An approximate analytical model of the transient response of a circular foil heat-flux gauge to a step change in incident radiant heat flux is described. This model demonstrates that the sensitivity and the transient response of the gauge are significantly affected by heat flow from the foil to the center lead wire. Results obtained for the transient response to a step change in incident radiant flux from the model agree closely with experimental results from several gauges with different sensitivities. Convolution of the present model and previous exponential models of the unit step response, with a transient incident flux represented by a terminal ramp, shows that the response predicted by a previous exponential model differs significantly from that predicted by the present model.Keywords
This publication has 3 references indexed in Scilit:
- Response of Finite-Thickness Gardon Heat-Flux SensorsJournal of Heat Transfer, 1972
- Response characteristics of thin foil heat flux sensorsAIAA Journal, 1969
- An Instrument for the Direct Measurement of Intense Thermal RadiationReview of Scientific Instruments, 1953