Improved signal quality indication for photoplethysmographic signals incorporating motion artifact detection
Authors:
- 1 August 2014
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- Vol. 2014 (1094687X) , 1872-1875
Abstract
Wearable monitoring systems have gained tremendous popularity in the health-care industry, opening new possibilities in diagnostic routines and medical treatments. Numerous hardware systems have been presented since, which allow for continuous acquisition of various biosignals like the ECG, PPG, EMG or EEG and which are suited for ambulatory settings. Unfortunately, these flexible systems are liable to motion artifacts and especially photoplethysmographic signals are seriously distorted when the patient is not at rest. A lot of work has been done to reduce artifacts and noise, ranging from simple filtering methods to very complex statistical approaches. With regard to the PPG, certain quality indices have been proposed to evaluate the signal conditions. As movements are the primary source of signal disturbances, the relation between the output of a signal quality estimator and acceleration data captured directly on the PPG sensor is focused in this work. It will be shown that typical motions can be detected on-line, thereby providing additional information which will significantly improve signal quality assessments.Keywords
This publication has 22 references indexed in Scilit:
- Dynamic time warping and machine learning for signal quality assessment of pulsatile signalsPhysiological Measurement, 2012
- Signal quality measures for pulse oximetry through waveform morphology analysisPhysiological Measurement, 2011
- A comparison of photoplethysmography and ECG recording to analyse heart rate variability in healthy subjectsJournal of Medical Engineering & Technology, 2009
- Motion Artifact Reduction in Photoplethysmographic Signals using Singular Value Decomposition2007
- Photoplethysmography and its application in clinical physiological measurementPhysiological Measurement, 2007
- Motion Artifact Reduction in Photoplethysmography Using Independent Component AnalysisIEEE Transactions on Biomedical Engineering, 2006
- Comparison of wavelet transformation and adaptive filtering in restoring artefact-induced time-related measurementBiomedical Signal Processing and Control, 2006
- Design of filter to reject motion artifact of pulse oximetryComputer Standards & Interfaces, 2004
- Determination of age-related increases in large artery stiffness by digital pulse contour analysisClinical Science, 2002
- Development of a neural network screening aid for diagnosing lower limb peripheral vascular disease from photoelectric plethysmography pulse waveformsPhysiological Measurement, 1993