PERFORMANCE EVALUATION OF DIAGNOSTIC RADIOLOGY DOSIMETERS IN CLINICAL AND CALIBRATION X-RAY BEAMS
Authors:
- Costantine John HourdakisCHCostantine John Hourdakis
- Argyro BoziariABArgyro Boziari
- Aggeliki ManetouAMAggeliki Manetou
- 1 May 2010
- journal article
- research article
- Published by Wolters Kluwer Health in Health Physics
- Vol. 98 (5) , 704-716
Abstract
Diagnostic radiology dosimeters should comply with International Electrotechnical Commission (IEC) 61674 standard in order to perform measurements with sufficient accuracy and reliability. The calibration of a dosimeter is performed under, and pertains to, reference conditions. However, in most cases, dosimeters are used for clinical measurements under non-reference conditions. The performance, in terms of accuracy of dose measurements, of six commercial diagnostic radiology dosimeters was tested at reference calibration and at clinical non-reference conditions. The results showed that all dosimeters being tested exhibited limits of variation within the +/-5% IEC limits. Depending on the detector's physical and operational properties, the dosimeters' energy dependence of response values varied from -4.7% to +4.2%. To address this variation of response, calibration at three radiation qualities (RQR 3, RQR 5, and RQR 9), at least, is recommended. Different irradiation conditions such as air kerma rate, x-ray tube design, x-ray system, and dosimeter operational modes affect the dosimeters' response by less than 3%. A dosimeter that complies with IEC standards and operates according to its specifications could be used at typical clinical irradiation conditions taking into account only corrections for the energy dependence of response. In this case, the error in dose accuracy is expected to be less than 3%.Keywords
This publication has 18 references indexed in Scilit:
- The effect of a compression paddle on energy response, calibration and measurement with mammographic dosimeters using ionization chambers and solid-state detectorsPhysics in Medicine & Biology, 2009
- The effect of spectra on calibration and measurement with mammographic ionization chambersMedical Physics, 2002
- Characteristics of radiation detectors for diagnostic radiologyApplied Radiation and Isotopes, 1999
- Ion recombination and polarity effect of ionization chambers in kilovoltage x‐ray exposure measurementsMedical Physics, 1998
- The Characterization of a Commercial Mosfet Dosimeter System for Use in Diagnostic X-RayHealth Physics, 1998
- General ion recombination for ionization chambers used under irradiation conditions relevant for diagnostic radiologyMedical Physics, 1995
- Spatial and Temporal Response Characteristics of Ionization Chambers Used in Diagnostic Radiology for Exposure Measurements and Quality ControlHealth Physics, 1994
- Requirements for Dosemeters Used in Diagnostic RadiologyRadiation Protection Dosimetry, 1992
- Recommendations on performance characteristics of diagnostic exposure meters: Report of AAPM Diagnostic X‐Ray Imaging Task Group No. 6Medical Physics, 1992
- Determination of Aion and Pion in the new AAPM radiotherapy dosimetry protocolMedical Physics, 1984