Photon time-of-flight (TOF) capability in positron emission tomography (PET) enables reconstructed image signal-to-noise ratio (SNR) improvement. With the coincidence time resolution (CTR) of 100 picosecond (ps), a five-fold SNR improvement can be achieved with a 40 cm diameter imaging subject, relative to a system without TOF capability. This 100 ps CTR can be achieved for a clinically relevant detector design (crystal element length ≥20 mm with reasonably high crystal packing fraction) using a side-readout PET detector configuration that enables 511 keV photon interaction depth-independent light collection efficiency and lower variance in scintillation photon transit time to the silicon photomultiplier (SiPM). In this study, we propose a new concept of TOF-PET detector to achieve high (95% accuracy in identifying interactions in the two adjacent 1.9 × 1.9 × 10 mm 3 crystal elements using the time-over-threshold technique. Based on our results, we can achieve excellent spatial and energy resolution in addition to ∼100 ps CTR with this novel detector design.