Dosimetric characteristics of LinaTech DMLC H multi leaf collimator: Monte Carlo simulation and experimental study
Open Access
- 6 March 2017
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 18 (2) , 113-124
Abstract
This study evaluated the basic dosimetric characteristics of a Dynamic Multi Leaf Collimator (DMLC) using a diode detector and film measurements for Intensity Modulated Radiation Therapy Quality Assurance (IMRT QA). The EGSnrc Monte Carlo (MC) simulation system was used for the determination of MLC characteristics. Radiation transmission and abutting leaf leakage relevant to the LinaTech DMLC H were measured using an EDGE detector and EBT3 film. In this study, the BEAMnrc simulation code was used for modeling. The head of Siemens PRIMUS linac (6 MV) with external DMLC H was entered into a BEAMnrc Monte Carlo model using practical dosimetry data. Leaf material density, as well as interleaf and abutting air gaps were determined according to the computed and measured dose profiles. The IMRT QA field was used to evaluate the dose distribution of the simulated DMLC H. According to measurements taken with the EDGE detector and film, the total average measured leakage was 1.60 ± 0.03% and 1.57 ± 0.05%, respectively. For these measurements, abutting leaf transmission was 54.35 ± 1.85% and 53.08 ± 2.05%, respectively. To adapt the simulated leaf dose profiles with measurements, leaf material density, interleaf and abutting air gaps were adjusted to 18 g/cm3, 0.008 cm and 0.108 cm, respectively. Thus, the total average leakage was estimated to be about 1.59 ± 0.02%. The step‐and‐shoot IMRT was implemented and 94% agreement was achieved between the film and MC, using 3%‐3 mm gamma criteria. The results of this study showed that the dosimetric characteristics of DMLC H satisfied international standards.Keywords
Funding Information
- Tehran University of Medical Sciences and Health Services (28202)
This publication has 31 references indexed in Scilit:
- Monte Carlo modeling and simulations of the High Definition (HD120) micro MLC and validation against measurements for a 6 MV beamMedical Physics, 2011
- Variability of a peripheral dose among various linac geometries for second cancer risk assessmentPhysics in Medicine & Biology, 2011
- Energy dependence and dose response of Gafchromic EBT2 film over a wide range of photon, electron, and proton beam energiesMedical Physics, 2010
- Impact of collimator leaf width and treatment technique on stereotactic radiosurgery and radiotherapy plans for intra- and extracranial lesionsRadiation Oncology, 2009
- Recommendations for implementing stereotactic radiotherapy in peripheral stage IA non-small cell lung cancer: report from the Quality Assurance Working Party of the randomised phase III ROSEL studyRadiation Oncology, 2009
- Influence of the linac design on intensity-modulated radiotherapy of head-and-neck plansPhysics in Medicine & Biology, 2006
- Weak energy dependence of EBT gafchromic film dose response in the 50kVp–10MVp X-ray rangeApplied Radiation and Isotopes, 2006
- Large efficiency improvements in BEAMnrc using directional bremsstrahlung splittingMedical Physics, 2004
- Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapyMedical Physics, 1998
- The Levenberg-Marquardt algorithm: Implementation and theory1978