Decoupling CE and ESI for a more robust interface with MS
- 26 March 2010
- journal article
- ce and-cec
- Published by Wiley in Electrophoresis
- Vol. 31 (7) , 1130-1137
Abstract
An interface for CE‐ESI‐MS that decouples both the electrical and the solution flow rate requirements of the separation and ionization processes is presented. The interface uses a tapered and beveled stainless steel hollow needle surrounding the separation capillary terminus so that the inside of the electrode acts as the CE outlet vial and the outside tip acts as the electrospray emitter. No capillary pre‐treatment is required, enabling the use of capillaries with any type of surface modification. A chemical modifier solution is introduced through a second capillary connected to the needle via a tee junction and can be used to improve the compatibility of the CE BGE with electrospray. The flow rate of modifier solution can be as low as 0.1 μL/min, much less than that in a typical sheath‐flow interface, thus minimizing dilution of the CE effluent in order to maximize sensitivity. The presence of the modifier solution also allows the use of neutral‐coated capillaries for protein analysis by CE‐MS without using an assisting pressure, despite the absence of EOF under these conditions. The interface is easily integrated into a commercial CE instrument, such that all operations can be carried out by the automated controls. Compared with a commercial sheath‐flow CE‐MS interface operating under optimized conditions, LODs for amino acids were, on average, improved fivefold.Keywords
This publication has 43 references indexed in Scilit:
- CE‐MS in metabolomicsElectrophoresis, 2009
- Metabolome analysis by capillary electrophoresis–mass spectrometryJournal of Chromatography A, 2007
- High resolution proteome/peptidome analysis of body fluids by capillary electrophoresis coupled with MSProteomics, 2006
- Pressure-assisted capillary electrophoresis mass spectrometry using combination of polarity reversion and electroosmotic flow for metabolomics anion analysisJournal of Bioscience and Bioengineering, 2006
- Combining capillary electrophoresis with mass spectrometry for applications in proteomicsElectrophoresis, 2005
- Recent applications of capillary electrophoresis‐electrospray ionisation‐mass spectrometry in drug analysisElectrophoresis, 2005
- Capillary electrophoresis – mass spectrometry: Survey on developments and applications 2003–2004Electrophoresis, 2005
- On‐line capillary electrophoresis‐mass spectrometry for the analysis of biomoleculesElectrophoresis, 2004
- Capillary Electrophoresis–Electrospray Ionization Mass Spectrometry of Amino Acids, Peptides, and Proteins2003
- Capillary electrophoresis – mass spectrometry: 15 years of developments and applicationsElectrophoresis, 2003