Validation of a Non-invasive Method Using Mechanical Extensometer for the Estimation of Intracranial Compliance by Repeated Measures Agreement Analysis
Preprint
- 19 February 2024
- preprint
- Published by Springer Nature in Research Square
Abstract
Validation of a Non-invasive Method Using Mechanical Extensometer for the Estimation of Intracranial Compliance Background Intracranial compliance refers to the relationship between changes in volume and the resultant changes in intracranial pressure (ICP). This study aimed to assess the validity of a non-invasive ICP waveform device for the estimation of compliance compared with invasive ICP monitoring employing three distinct methods. Methods We conducted a retrospective analysis of ICP waveform morphology recorded through both invasive (external ventricular drain) and non-invasive (mechanical extensometer) methods in adult patients with acute brain injury admitted to the Neurointensive Care Unit at the Cleveland Clinic between 8/2021 to 8/2022. Compliance was calculated as the amplitude of the fundamental component of cerebral arterial blood volume (estimated with concurrent Transcranial Doppler [TCD] recordings), divided by the amplitude of the fundamental component of the invasive and non-invasive ICP waveforms. Subsequently, we assessed the agreement between invasive ICC and non-invasive ICC by repeated measures correlation coefficient analysis using 3 methods: TCD-derived, P2/P1 ratio, time-to-peak (TTP). Linear mixed-effects model was used to compute the concordance correlation coefficient (CCC), total deviation index (TDI), and coefficient of individual agreement (CIA). Coverage probability plot (CPP) was calculated to estimate the percent of observations within different cut points for each of the three methods. Results 21 patients were identified. Repeated measures correlation analysis showed a strong correlation R=0.982 (95% CI [0.980, 0.984], p<0.0001) between log-transformed non-invasive and invasive compliance. Agreement statistics for TCD, P2/P1 ratio, and TTP demonstrated that although the CCC was highest for log(TCD) values, TTP and P2:P1 ratio measures had better agreement with TDI and CPP analyses. Conclusion Repeated measures correlations suggest that ICP waveform analyses may offer a more accurate estimate of compliance than TCD-derived methods for non-invasive ICP monitoring. Further validations studies are warranted to confidently establish this method as a reliable indicator of intracranial compliance.Keywords
All Related Versions
- Version 1, 2024-02-19, Research Square
- Published version: Neurocritical Care, 42 (2), 465.
This publication has 19 references indexed in Scilit:
- Repeated Measures CorrelationFrontiers in Psychology, 2017
- Transcranial Doppler Monitoring of Intracranial Pressure Plateau WavesNeurocritical Care, 2016
- Monro-Kellie 2.0: The dynamic vascular and venous pathophysiological components of intracranial pressureJournal of Cerebral Blood Flow & Metabolism, 2016
- Correlation and agreement: overview and clarification of competing concepts and measures2016
- Non-invasive Monitoring of Intracranial Pressure Using Transcranial Doppler Ultrasonography: Is It Possible?Neurocritical Care, 2016
- Validation of a New Noninvasive Intracranial Pressure Monitoring Method by Direct Comparison with an Invasive Technique2016
- Validation of a New Minimally Invasive Intracranial Pressure Monitoring Method by Direct Comparison with an Invasive Technique2016
- Model-based Indices Describing Cerebrovascular DynamicsNeurocritical Care, 2013
- The monitoring of relative changes in compartmental compliances of brainPhysiological Measurement, 2009
- Cerebral Vascular Mean Transit Time in Healthy Humans: A Comparative Study with PET and Dynamic Susceptibility Contrast-Enhanced MRIJournal of Cerebral Blood Flow & Metabolism, 2007