Critical flow venturi nozzles (CFVNs) are expected to play a key role for metrology of gases in the future due to their affordability, reliability and flexibility. While their performance with several inert gases is well documented, there is a lack of experimental data when used with hydrogen. The objective of this work is to estimate the relevance of calibrating CFVNs with air as a surrogate gas for hydrogen, especially at transitional and turbulent flow regimes. Two nozzles with diameters 1 mm and 2 mm are calibrated with both hydrogen gas and air using a primary standard based on a pressure–Volume–Temperature–time system. The results show that the discharge coefficients of the CFVNs for hydrogen and air are within the expanded uncertainty of ± 0.3   % ( k = 2) of the laboratory.