The statistical uncertainty of the Heydemann correction: a practical limit of optical quadrature homodyne interferometry
- 1 July 2015
- journal article
- Published by IOP Publishing in Measurement Science and Technology
- Vol. 26 (8)
Abstract
Although the Heydemann correction is widely used to demodulate the phase of quadrature homodyne interferometer and encoder signals, the related measurement uncertainty is considered only in a few publications. The statistical uncertainty of the phase, which is the contribution of the high-frequency noise in the input signals, due to the use of a simplified fitting approach in the Heydemann correction, cannot be determined exactly. Therefore, neglecting the correlation between the input signals and the fit parameters, an approximation is provided and verified using Monte Carlo simulations and an iterative fitting method that allows exact determination of the statistical uncertainty of the phase. For the high-quality signals encountered in dimensional metrology applications, only minor differences of a few percent in the uncertainty estimates are observed. Furthermore, for cases in which a number of points used in the fit is sufficiently large, a new, simple, analytic expression for the statistical uncertainty of the phase is derived. It represents a practical limit of optical quadrature displacement interferometry, which already has been reached experimentally. To achieve an expanded measurement uncertainty of 10 pm a signal-to-noise greater than 10 4 has to be realized experimentally.Keywords
This publication has 21 references indexed in Scilit:
- Ellipse fitting for interferometry Part 1: static methodsJournal of the Optical Society of America A, 2014
- A high performance one-dimensional homodyne encoder and the proof of principle of a novel two-dimensional homodyne encoderPrecision Engineering, 2013
- Interference signal demodulation for nanopositioning and nanomeasuring machinesMeasurement Science and Technology, 2012
- Comparison of the performance of the next generation of optical interferometersMetrologia, 2012
- Enhanced ellipse fitting in a two-detector homodyne quadrature laser interferometerMeasurement Science and Technology, 2011
- Linearisation of counting interferometers with 0.1 nm precisionInternational Journal of Nanotechnology, 2007
- Simple, real-time method for removing the cyclic error of a homodyne interferometer with a quadrature detector systemApplied Optics, 2005
- Verification of the sub-nanometric capability of an NPL differential plane mirror interferometer with a capacitance probeMeasurement Science and Technology, 1998
- Correction of nonlinearity in one-frequency optical interferometryMeasurement Science and Technology, 1996
- Determination and correction of quadrature fringe measurement errors in interferometersApplied Optics, 1981