The power of three: Atrial rate limit power‐on‐reset — A non‐electromagnetic cause of pacemaker‐induced extracardiac stimulation
- 13 June 2023
- journal article
- editorial
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 47 (1) , 113-116
Abstract
BackgroundPower-on reset (PoR) is most commonly due to electromagnetic interference. Full PoR results in a switch to an inhibited mode (VVI) pacing and resets pacing outputs to maximal unipolar settings, leading to extracardiac stimulation. MethodsWe present a case of PoR occurrence in the absence of electromagnetic interference, resulting in pectoral stimulation triggered by violation of the atrial rate limit. ConclusionsIt is useful for clinicians to recognizethe occurrence of PoR in the setting of atrial limit violation andthe appropriate management in such circumstances.Keywords
This publication has 8 references indexed in Scilit:
- 2018 ACC/AHA/HRS Guideline on the Evaluation and Management of Patients With Bradycardia and Cardiac Conduction DelayJACC, 2019
- Cause of the “power-on reset” phenomenon other than electric magnetic interference in a patient with a pacemakerIndian Pacing and Electrophysiology Journal, 2018
- “Power-on resets” in cardiac implantable electronic devices during magnetic resonance imagingHeart Rhythm, 2015
- Effects of External Electrical and Magnetic Fields on Pacemakers and DefibrillatorsCirculation, 2013
- Cosmic radiation induced software electrical resets in ICDs during air travelHeart Rhythm, 2008
- Safety of Magnetic Resonance Imaging in Patients With Cardiovascular DevicesCirculation, 2007
- Runaway pacemaker: A forgotten phenomenon?EP Europace, 2005
- Case Report: Complex Arrhythmia Induced by Noncompetitive Atrial Pacing Algorithm of DDDR PacemakerJournal of Interventional Cardiac Electrophysiology, 2001