Time Discrimination Techniques Using Artificial Neural Networks for Positron Emission Tomography
Authors:
- 16 June 2009
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 56 (3) , 588-595
Abstract
Relevant information in positron emission tomography is currently being obtained mostly by analog signal-processing methods. New digital PET scanner architectures are now becoming available, which offer greater flexibility and easier reconfiguration capability as compared to previous PET designs. Moreover, new strategies can be devised to extract more information with better accuracy from the digitized detector signals. Trained artificial neural networks (ANN) have been investigated to improve coincidence timing resolution with different types of Avalanche PhotoDiode (APD)-based detectors. The signal at the output of a charge-sensitive preamplifier was digitized with an off-the-shelf, free-running 100-MHz, 8-bit analog-to-digital converter and time discrimination was performed with ANNs implemented in field-programmable gate array (FPGA). Results show that ANNs can be particularly efficient with slow and low light output scintillators, such as BGO, but less so with faster luminous crystals, such as LSO. In reference to a fast PMT-plastic detector, a time resolution of 6.5 ns was achieved with a BGO-APD detector. With LSO, the ANN was found to be competitive with other digital techniques developed in previous works. ANNs implemented in FPGAs provide a fast and flexible circuit that can be easily reconfigured to accommodate various detectors under different signal/noise conditions.Keywords
This publication has 12 references indexed in Scilit:
- The Hardware and Signal Processing Architecture of LabPET™, a Small Animal APD-Based Digital PET ScannerIEEE Transactions on Nuclear Science, 2009
- Performance evaluation of the LabPET™ APD-based digital PET scanner2007
- Real time digital signal processing implementation for an APD-based PET scanner with phoswich detectorsIEEE Transactions on Nuclear Science, 2006
- Architecture of a dual-modality, high-resolution, fully digital positron emission tomography/computed tomography (PET/CT) scanner for small animal imagingIEEE Transactions on Nuclear Science, 2005
- Time determination of BGO-APD detectors by digital signal processing for positron emission tomography2003
- A PET system with free running ADCsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2002
- Performance analysis of phoswich/APD detectors and low-noise CMOS preamplifiers for high-resolution PET systemsIEEE Transactions on Nuclear Science, 2001
- Adaptive digital spectroscopy in programmable logicIEEE Transactions on Nuclear Science, 2000
- A power-efficient, low-noise, wideband, integrated CMOS preamplifier for LSO/APD PET systemsIEEE Transactions on Nuclear Science, 2000
- High resolution positron emission tomography with a prototype camera based on solid state scintillation detectorsIEEE Transactions on Nuclear Science, 1990