Abstract
Objectives An ongoing prospective study to investigate failing metal-on-metal hip prostheses was commenced at our centre in 2008. We report on the results of the analysis of the first consecutive 126 failed mated total hip prostheses from a single manufacturer. Methods Analysis was carried out using highly accurate coordinate measuring to calculate volumetric and linear rates of the articular bearing surfaces and also the surfaces of the taper junctions. The relationship between taper wear rates and a number of variables, including bearing diameter and orientation of the acetabular component, was investigated. Results The measured rates of wear and distribution of material loss from the taper surfaces appeared to show that the primary factor leading to taper failure is the increased lever arm acting on this junction in contemporary large-diameter metal-on-metal hip replacements. Conclusions Our analysis suggests that varus stems, laterally engaging taper systems and larger head diameters all contribute to taper failure.Keywords
This publication has 23 references indexed in Scilit:
- Accelerating failure rate of the ASR total hip replacementThe Journal of Bone and Joint Surgery. British volume, 2011
- Large Diameter Femoral Heads Impose Significant Alterations on the Strains Developed on Femoral Component and Bone: A Finite Element AnalysisThe Open Orthopaedics Journal, 2011
- The John Charnley Award: Metal-on-Metal Hip Resurfacing versus Large-diameter Head Metal-on-Metal Total Hip ArthroplastyClinical Orthopaedics and Related Research, 2010
- Revision of metal-on-metal resurfacing arthroplasty of the hipThe Journal of Bone and Joint Surgery. British volume, 2008
- Pseudotumours associated with metal-on-metal hip resurfacingsThe Journal of Bone and Joint Surgery. British volume, 2008
- In vitro comparison of frictional torque and torsional resistance of aged conventional gamma-in-nitrogen sterilized polyethylene versus aged highly crosslinked polyethylene articulating against head sizes larger than 32 mmActa Orthopaedica, 2006
- Excellent results of HA coating on a grit-blasted stem: 245 patients followed for 8-12 yearsActa Orthopaedica, 2003
- In vitro corrosion testing of modular hip tapersJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2002
- Effect of manufacturing tolerances on the micromotion at the Morse taper interface in modular hip implants using the finite element techniqueBiomaterials, 1996