Abstract
OBJECTIVES: Although commercial iliac branch devices offer a new and valid endovascular approach to treating iliac aneurysm and effectively preserve antegrade flow of the internal iliac artery, their use may not be suited for all types of challenging anatomy, especially isolated common iliac artery aneurysm. Our custom-made iliac bifurcation device has a unique design and excludes both combined and isolated iliac branch aneurysm. This study validated the efficacy and safety of the custom device by comparing clinical outcomes between groups receiving commercial and custom devices. METHODS: Data of consecutive patients receiving iliac bifurcation device implantation for iliac aneurysm with or without concomitant endovascular repair for abdominal aortic aneurysm from January 2010 to May 2019 were reviewed. RESULTS: Iliac bifurcation device implantation with or without concomitant abdominal aortic aneurysm stent grafting was completed in 46 patients (commercial, n = 35; custom, n = 11). No significant differences were observed regarding postoperative complications, occlusion or endoleak. Comparisons of primary (80.8% vs 85.7%, P = 0.88) and secondary (86.5% vs 85.7%, P = 0.85) patency and freedom from reintervention (88.2% vs 100%, P = 0.33), all-cause mortality (78.6% vs 100%, P = 0.25) and aneurysm-related mortality (100% vs 100%, P = 1.00) also indicated no differences at a 5-year surveillance point. Furthermore, the iliac aneurysms of the groups displayed similar shrinkage 1 year after procedures. CONCLUSIONS: For iliac aneurysm, the novel custom-made iliac bifurcation device is an adaptable design not inferior to commercial devices with regard to postoperative complications, bridge occlusion, endoleak and short-term aneurysm remodelling. It provides an alternative for treatment, particularly when certain anatomic challenges are present. Clinical trial registration: 2018-07-050BC, 2017-01-023ACF.Keywords
Funding Information
- National Science Council (MOST105-2314-B-010-048-MY3)
- Taipei Medical University (TMU108-F-005, TMU108-AE1-B18)
- Wan Fang Hospital (109-wf-eva-02, 109-wf-eva-07, 109TMU-WFH-19)
This publication has 18 references indexed in Scilit:
- Effect of a Flared Renal Stent on the Performance of Fenestrated Stent-Grafts at Rest and Exercise ConditionsJournal of Endovascular Therapy, 2016
- Comparison between endovascular repair and open surgery for isolated iliac artery aneurysmsSurgery Today, 2014
- Outcomes of open and endovascular repair for ruptured and nonruptured internal iliac artery aneurysmsJournal of Vascular Surgery, 2014
- Survey of management of common iliac artery aneurysms by members of the Vascular Society of Great Britain and IrelandThe Annals of The Royal College of Surgeons of England, 2014
- The sandwich technique to treat complex aortoiliac or isolated iliac aneurysms: Results of midterm follow-upJournal of Vascular Surgery, 2013
- Long-term Results of Iliac Aneurysm Repair with Iliac Branched Endograft: A 5-Year Experience on 100 Consecutive CasesEuropean Journal of Vascular and Endovascular Surgery, 2012
- Fenestrated and Branched Stent GraftsPerspectives in Vascular Surgery and Endovascular Therapy, 2008
- Solitary aneurysms of the iliac arterial system: An estimate of their frequency of occurrenceJournal of Vascular Surgery, 1989
- Solitary aneurysms of the iliac arterial system: An estimate of their frequency of occurrenceJournal of Vascular Surgery, 1989
- Isolated Aneurysms of the Iliac ArteryArchives of Surgery, 1978