3D model-based super-resolution motion-corrected cardiac T1 mapping
Authors:
- Simone HufnagelSHSimone Hufnagel
- Selma MetznerSMSelma Metzner
- Kirsten Miriam KerkeringKKKirsten Miriam Kerkering
- Christoph Stefan AignerCAChristoph Stefan Aigner
- Andreas KoflerAKAndreas Kofler
- Jeanette Schulz-MengerJSJeanette Schulz-Menger
- Tobias SchaeffterTSTobias Schaeffter
- Christoph KolbitschCKChristoph Kolbitsch
Open Access
- 9 December 2022
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 67 (24) , 245008
Abstract
Objective. To provide 3D high-resolution cardiac T1 maps using model-based super-resolution reconstruction (SRR). Approach. Due to signal-to-noise ratio limitations and the motion of the heart during imaging, often 2D T1 maps with only low through-plane resolution (i.e. slice thickness of 6–8 mm) can be obtained. Here, a model-based SRR approach is presented, which combines multiple stacks of 2D acquisitions with 6–8 mm slice thickness and generates 3D high-resolution T1 maps with a slice thickness of 1.5–2 mm. Every stack was acquired in a different breath hold (BH) and any misalignment between BH was corrected retrospectively. The novelty of the proposed approach is the BH correction and the application of model-based SRR on cardiac T1 Mapping. The proposed approach was evaluated in numerical simulations and phantom experiments and demonstrated in four healthy subjects. Main results. Alignment of BH states was essential for SRR even in healthy volunteers. In simulations, respiratory motion could be estimated with an RMS error of 0.18 ± 0.28 mm. SRR improved the visualization of small structures. High accuracy and precision (average standard deviation of 69.62 ms) of the T1 values was ensured by SRR while the detectability of small structures increased by 40%. Significance. The proposed SRR approach provided T1 maps with high in-plane and high through-plane resolution (1.3 × 1.3 × 1.5–2 mm3). The approach led to improvements in the visualization of small structures and precise T1 values.Keywords
Funding Information
- Deutsche Forschungsgemeinschaft (GRK2260)
- EMPIR Programme (18HLT05)
This publication has 64 references indexed in Scilit:
- Myocardial T1 and T2 mapping at 3 T: reference values, influencing factors and implicationsJournal of Cardiovascular Magnetic Resonance, 2013
- Reconstruction of fetal brain MRI with intensity matching and complete outlier removalMedical Image Analysis, 2012
- Super‐resolution in magnetic resonance imaging: A reviewConcepts in Magnetic Resonance Part A, 2012
- T2 quantification for improved detection of myocardial edemaJournal of Cardiovascular Magnetic Resonance, 2009
- A New Alternating Minimization Algorithm for Total Variation Image ReconstructionSIAM Journal on Imaging Sciences, 2008
- Undersampled radial MRI with multiple coils. Iterative image reconstruction using a total variation constraintMagnetic Resonance in Medicine, 2007
- “Soap‐Bubble” visualization and quantitative analysis of 3D coronary magnetic resonance angiogramsMagnetic Resonance in Medicine, 2002
- Standardized Myocardial Segmentation and Nomenclature for Tomographic Imaging of the HeartCirculation, 2002
- Nonrigid registration using free-form deformations: application to breast MR imagesIEEE Transactions on Medical Imaging, 1999
- Parameter relations for the Shinnar-Le Roux selective excitation pulse design algorithm (NMR imaging)IEEE Transactions on Medical Imaging, 1991