Monte Carlo Simulation of Low-Count Signals in Time-of-Flight Mass Spectrometry and Its Application to Single-Particle Detection
Open Access
- 21 September 2018
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 90 (20) , 11847-11855
Abstract
Many modern time-of-flight mass spectrometry (TOFMS) instruments use fast analog-to-digital conversion (ADC) with high-speed digitizers to record mass spectra with extended dynamic range (compared to time-to-digital conversion). The extended dynamic range offered by ADC detection is critical for accurate measurement of transient events. However, the use of fast ADC also increases the variance of the measurements by sampling the gain statistics of electron multipliers (EMs) used for detection. The influence of gain statistics on the shape of TOF signal distributions is especially pronounced at low count rates and is a major contributor to measurement variance. Here, we use Monte Carlo methods to simulate low-ion-count TOFMS signals as a function of Poisson statistics and the measured pulse-height distribution (PHD) of the EM-detection system. We find that a compound Poisson distribution calculated via Monte Carlo simulation effectively describes the shape of measured TOFMS signals. Additionally, we apply Monte Carlo simulation results to single-particle inductively coupled plasma TOFMS (sp-ICP-TOFMS) analysis. We demonstrate that subtraction of modelled TOFMS signals can be used to quantitatively uncover particle-signal distributions buried beneath dissolved-signal backgrounds. Based on simulated signal distributions, we also calculate new critical values (LC) that are used as decision thresholds for the detection of discrete particles. This new detection criterion better accounts for the shape of dissolved signal distributions and therefore provides more robust identification of single particles with ICP-TOFMS.Keywords
Funding Information
- Schweizerischer Nationalfonds zur Förderung der Wissenschaftlichen Forschung (PZ00P2_174061, 200021 162870/1)
This publication has 48 references indexed in Scilit:
- Time-of-flight mass spectrometry (TOFMS): From niche to mainstreamInternational Journal of Mass Spectrometry, 2015
- Simultaneous Mass Quantification of Nanoparticles of Different Composition in a Mixture by Microdroplet Generator-ICPTOFMSAnalytical Chemistry, 2014
- History of mass measurements in time-of-flight mass analyzersInternational Journal of Mass Spectrometry, 2013
- Mass spectrometry of atmospheric aerosols—Recent developments and applications. Part II: On‐line mass spectrometry techniquesMass Spectrometry Reviews, 2011
- Chemical and microphysical characterization of ambient aerosols with the aerodyne aerosol mass spectrometerMass Spectrometry Reviews, 2007
- Multi‐turn time‐of‐flight mass spectrometers with electrostatic sectorsJournal of Mass Spectrometry, 2003
- Concepts and Preliminary Observations on the Triple-Dimensional Analysis of Complex Volatile Samples by Using GC×GC−TOFMSAnalytical Chemistry, 2001
- Orthogonal acceleration time-of-flight mass spectrometryMass Spectrometry Reviews, 2000
- Three-Dimensional Ion Mobility/TOFMS Analysis of Electrosprayed BiomoleculesAnalytical Chemistry, 1998
- Molecular Imaging of Biological Samples: Localization of Peptides and Proteins Using MALDI-TOF MSAnalytical Chemistry, 1997