Reliability of vessel diameter measurements with a retinal oximeter
- 16 April 2011
- journal article
- clinical trial
- Published by Springer Nature in Graefe's Archive for Clinical and Experimental Ophthalmology
- Vol. 249 (9) , 1311-1317
Abstract
The purpose of this study was to test the reliability of vessel diameter measurements with a newly developed retinal oximeter. Twelve healthy individuals participated in the study. Retinal images were taken with the Oxymap Retinal Oximeter. Diameters of retinal vessels were measured automatically with the Oxymap Analyzer software. Repeated measurements on the same vessel segments were compared. The automatic measurements were also compared with semi-automatic measurements with a plug-in for the ImageJ software. Variance coefficient (standard deviation/mean) from the repeatability test was 2.8 and 4.0% for first- and second-degree venules and 3.5 and 5.4% for first- and second-degree arterioles, respectively. ImageJ measured larger mean diameters than Oxymap Analyzer in all cases. Means of differences were 5.1 ± 2.2, 2.9 ± 1.3, and 2.7 ± 1.6 pixels for first-, second-, and third-degree venules and 3.1 ± 1.2, 2.7 ± 0.9, and 2.9 ± 1.4 pixels for first-, second-, and third-degree arterioles. Vessel diameter measurements with the oximeter are repeatable and comparison with an established method demonstrates a relatively stable offset where the standard deviation of the difference is rather small. Different definitions of vessel borders may be the cause of this difference.Keywords
This publication has 53 references indexed in Scilit:
- RETINAL VESSEL ANALYZER MEASUREMENTS OF THE EFFECT OF PANRETINAL PHOTOCOAGULATION ON THE RETINAL ARTERIOLAR DIAMETER IN DIABETIC RETINOPATHYRetina, 2010
- Correlation of Flicker-Induced and Flow-Mediated Vasodilatation in Patients With Endothelial Dysfunction and Healthy VolunteersDiabetes Care, 2009
- Measurement of total blood flow in the normal human retina using Doppler Fourier-domain optical coherence tomographyBritish Journal of Ophthalmology, 2009
- Computerized Analysis of Retinal Vessel Width and Tortuosity in Premature InfantsInvestigative Ophthalmology & Visual Science, 2008
- Automatic Retinal OximetryInvestigative Ophthalmology & Visual Science, 2006
- Retinal vessel diameter and open-angle glaucomaOphthalmology, 2005
- Variation associated with measurement of retinal vessel diameters at different points in the pulse cycleBritish Journal of Ophthalmology, 2004
- Retinal Vessel Diameters and Their Associations with Age and Blood PressureInvestigative Ophthalmology & Visual Science, 2003
- Acute effects of oxygen and carbon dioxide on retinal vascular network geometry in hypertensive and normotensive subjectsClinical Science, 2000
- Evaluation of the Zeiss retinal vessel analyserBritish Journal of Ophthalmology, 2000