First experience with single-source, dual-energy CCTA for monochromatic stent imaging
- 18 December 2014
- journal article
- research article
- Published by Oxford University Press (OUP) in European Heart Journal - Cardiovascular Imaging
- Vol. 16 (5) , 507-512
Abstract
Aims Single-source, dual-energy coronary computed tomography angiography (CCTA) with monochromatic image reconstruction allows significant noise reduction. The aim of the study was to evaluate the impact of monochromatic CCTA image reconstruction on coronary stent imaging, as the latter is known to be affected by artefacts from highly attenuating strut material resulting in artificial luminal narrowing. Methods and results Twenty-one patients with 62 stents underwent invasive coronary angiography and single-source, dual-energy CCTA after stent implantation. Standard polychromatic images as well as eight monochromatic series (50, 60, 70, 80, 90, 100, 120, and 140 keV) were reconstructed for each CCTA. Signal and noise were measured within the stent lumen and in the aortic root. Mean in-stent luminal diameter was assessed in all CCTA reconstructions and compared with quantitative invasive coronary angiography (QCA). Luminal attenuation was higher in the stent than in the aortic root throughout all monochromatic reconstructions ( P < 0.001). An increase in monochromatic energy was associated with a decrease in luminal attenuation values ( P < 0.001). The mean in-stent luminal diameter underestimation by monochromatic CCTA compared with QCA was 90% at low monochromatic energy (50 keV) and improved to 37% at high monochromatic (140 keV) reconstruction while stent diameter was underestimated by 39% with standard CCTA. Conclusion Monochromatic CCTA can be used reliably in patients with coronary stents. However, reconstructions with energies below 80 keV are not recommended as the blooming artefacts are most pronounced at such low energies, resulting in up to 90% stent diameter underestimation.Keywords
This publication has 27 references indexed in Scilit:
- Edge Vascular Response After Percutaneous Coronary InterventionJACC: Cardiovascular Interventions, 2013
- Incremental Prognostic Value of Cardiac Computed Tomography in Coronary Artery Disease Using CONFIRMCirculation: Cardiovascular Imaging, 2011
- Prognostic performance of low-dose coronary CT angiography with prospective ECG triggeringHeart, 2011
- Meta-Analysis of Diagnostic Efficacy of 64-Slice Computed Tomography in the Evaluation of Coronary In-Stent RestenosisThe American Journal of Cardiology, 2009
- Comparison of Feasibility and Diagnostic Accuracy of 64-Slice Multidetector Computed Tomographic Coronary Angiography Versus Invasive Coronary Angiography Versus Intravascular Ultrasound for Evaluation of In-Stent RestenosisThe American Journal of Cardiology, 2009
- Noninvasive Assessment of Coronary In-Stent Restenosis by Dual-Source Computed TomographyThe American Journal of Cardiology, 2009
- Accuracy of MDCT in Assessing the Degree of Stenosis Caused by Calcified Coronary Artery PlaquesAmerican Journal of Roentgenology, 2008
- Feasibility of assessment of coronary stent patency using 16-slice computed tomographyThe American Journal of Cardiology, 2004
- Multislice spiral computed tomography for the detection of coronary stent restenosis and patencyInternational Journal of Cardiology, 2003
- Imaging of coronary artery stents using multislice computed tomography: in vitro evaluationEuropean Radiology, 2003