Integration of fixed-frequency and FM-CW (frequency-modulated continuous-wave) reflectometers for coincident turbulence measurements on LTX-β (Lithium Tokamak eXperiment-β)
- 1 November 2024
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 95 (11)
Abstract
The fixed-frequency and frequency-modulated continuous-wave (FM-CW) reflectometers on LTX-β (Lithium Tokamak eXperiment-β) have been configured to use the same transmission lines and antenna arrays for coincident views of the core and edge plasma. The fixed-frequency channels (13.1–20.5 and 20–40 GHz, tunable between discharges) provide time-resolved measurements of density fluctuations, while the FM-CW channels (13.1–20.2 and 19.5–33.5 GHz) measure the density profile and fluctuations, with high spatial resolution and a sampling rate determined by the frequency sweep interval (5 μs). Data from both reflectometers are synchronously acquired to simultaneously leverage the wide bandwidth and high spatial resolution of the respective systems. Experiments showed that mutual crosstalk interference is momentary and does not diminish the capability of either system. Spectral analysis indicated broad power spectra (several hundreds of kHz) and suggests that the signals from the FM-CW system are consistent with under-sampled fixed-frequency signals. Radial correlations were explored using data from the two reflectometers, as well as from the FM-CW system alone. The core channels showed high levels of agreement between these two comparisons, suggesting that the data from the reflectometers are interchangeable for statistical estimates. For the edge channels, comparisons using data from the FM-CW reflectometer alone showed significant decorrelation due to time lag caused by the finite frequency up-sweep duration. Alternatively, this effect is eliminated when cross-correlating data from the different reflectometers. These results highlight the advantages of operating the fixed-frequency and FM-CW reflectometers in this manner, where the combined system can overcome the limitations of each separate system.Keywords
Funding Information
- U.S. Department of Energy (DE-AC02-09CH11466, DE- SC0023481)
This publication has 16 references indexed in Scilit:
- Handbook of Signal Processing Systems2019
- Millimeter-wave interferometry and far-forward scattering for density fluctuation measurements on LTX-βReview of Scientific Instruments, 2018
- 1 μs broadband frequency sweeping reflectometry for plasma density and fluctuation profile measurementsReview of Scientific Instruments, 2017
- A frequency-modulated continuous-wave reflectometer for the Lithium Tokamak ExperimentReview of Scientific Instruments, 2017
- Compatibility of lithium plasma-facing surfaces with high edge temperatures in the Lithium Tokamak ExperimentPhysics of Plasmas, 2017
- A Ka-band tunable direct-conversion correlation reflectometer for NSTXReview of Scientific Instruments, 2010
- Comparison of turbulence measurements from DIII-D low-mode and high-performance plasmas to turbulence simulations and modelsPhysics of Plasmas, 2002
- The UCLA frequency tunable correlation reflectometer system on DIII-DReview of Scientific Instruments, 1992
- ATF two-frequency correlation reflectometerReview of Scientific Instruments, 1990
- The real height analysis of ionograms: A generalized formulationRadio Science, 1988