Application of contrast-to-noise ratio in optimizing beam quality for digital chest radiography: comparison of experimental measurements and theoretical simulations
Authors:
- P DoylePDP Doyle
- C J MartinCMC J Martin
- D GentleDGD Gentle
- 24 May 2006
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 51 (11) , 2953-2970
Abstract
The contrast-to-noise ratio (CNR) has been employed in optimizing beam quality for imaging a chest phantom using digital radiography. The relationship between CNR and tube potential has been studied for regions of different attenuations representing the lung, heart and abdomen, and a figure of merit (FOM) incorporating effective dose has been calculated to enable dose performance to be included. Direct measurements of imaging performance have been compared with simulations based on a model representing object attenuations. The study has shown reasonable agreement between measurements of CNR and calculated values. The CNR values in the lung and heart regions are higher at 60-80 kV, while those for the abdomen are higher at 90-110 kV. Incorporating a 0.2 mm copper filter has minimal effect on image quality and the FOM is higher because of the reduction in dose. For imaging the heart and abdomen, performance was improved through use of a technique to remove scatter, with an air-gap technique giving a higher FOM because of the lower dose. The CNR and FOM can provide useful quantities for evaluating imaging performance to optimize beam quality for different imaging tasks.Keywords
This publication has 15 references indexed in Scilit:
- Dose–image quality optimisation in digital chest radiographyRadiation Protection Dosimetry, 2005
- Radiation Exposure and Image Quality in X-Ray Diagnostic Radiology2004
- Chest Radiography: Optimization of X-ray Spectrum for Cesium Iodide–Amorphous Silicon Flat-Panel DetectorRadiology, 2003
- The influence of different technique factors on image quality of chest radiographs as evaluated by modified CEC image quality criteriaBritish Journal of Radiology, 2002
- Towards image quality, beam energy and effective dose optimisation in digital thoracic radiographyEuropean Radiology, 2001
- Anthropomorphic versus geometric chest phantoms: A comparison of scatter propertiesMedical Physics, 2000
- Quality control phantom for digital chest radiography.Radiology, 1997
- Digital chest radiography with photostimulable storage phosphors: signal-to-noise ratio as a function of kilovoltage with matched exposure risk.Radiology, 1993
- Physical model for the contrast sensitivity of the human eye1992
- Threshold perception performance with computed and screen-film radiography: implications for chest radiography.Radiology, 1992