Bone Suture Anchor Fixation in the Lower Extremity: A Review of Insertion Principles and a Comparative Biomechanical Evaluation
- 1 July 2005
- journal article
- research article
- Published by SAGE Publications in Foot & Ankle International
- Vol. 26 (7) , 516-519
Abstract
Background: Suture anchors have been developed for the fixation of ligaments, capsules, or tendons to bone. These devices have led to improved fixation, smaller incisions, earlier limb mobility, and improved outcomes. They were originally developed for use in shoulder reconstructions but are now used in almost all extremities. In the lower leg they are used in the tibia, the talus, the calcaneus, tarsal bones, and phalanges. Nevertheless, techniques for insertion and mechanisms of failure are not well described. Methods: Five suture anchors were studied to determine the pullout strength in four distal cadaver femurs and four proximal cadaver tibias from 55- and 62-year-old males. Eight hundred ninety Newton line was used, testing the anchors to failure with an Instron testing device (Instron, Norwood, MA). The anchor devices were inserted randomly and tested blindly (12 tests per anchor device, 60 tests in all). Results: Two anchors in each group tested failed at low loads. Both types of plastic anchors had failures at the eyelet. Average pullout strength varied from 85.4 to 185.6 N. Conclusions: Insertion techniques are specific for each device, and they must be followed for optimal fixation. In this study, in all five groups of anchors tested two of the 12 anchors in each group failed with minimal force. On the basis of this finding we recommend that, if suture anchor fixation is necessary, at least two anchors should be used. Since there appears to be a percentage of failure in all devices, the second anchor can serve as a backup. It is imperative that surgeons be familiar with the insertion techniques of each device before use.Keywords
This publication has 13 references indexed in Scilit:
- The effect of suture anchor design and orientation on suture abrasion: An in vitro studyArthroscopy, 2003
- Bone suture anchors versus the pullout button for repair of distal profundus tendon injuries: A comparison of strength in human cadaveric handsThe Journal of Hand Surgery, 2001
- Biological and mechanical characteristics of the interface between a new swelling anchor and boneJournal of Orthopaedic Research, 2000
- Suture Anchors—Update 1999Arthroscopy, 1999
- The relationship of suture anchor failure and bone density to proximal humerus location: A cadaveric studyArthroscopy, 1997
- Internal fixation strength of suture anchors—Update 1997Arthroscopy, 1997
- The pullout force for Mitek mini and micro suture anchor systems in human mandibular condylesJournal of Oral and Maxillofacial Surgery, 1997
- Suture anchor strength revisitedArthroscopy, 1996
- Comparative Evaluation of Bone Suture Anchor to Bone Tunnel Fixation of Tibialis Anterior Tendon in Cadaveric Cuboid Bone: A Biomechanical InvestigationFoot & Ankle International, 1995
- The deadman theory of suture anchors: observations along a South Texas fence lineArthroscopy, 1995