Abstract
A lincosamide antibiotic injectable formulation contains clindamycin phosphate (CMN), benzyl alcohol (BAL), and Ethylenediaminetetraacetic acid (EDTA), all the contents were qualitatively and quantitatively determined by a single green liquid chromatography method. The three components were separated with Phenomenex prodigy C8 (250 mm x 4.6 mm, 5µm) HPLC column, flow rate 1.1 mL/min, injection volume 30 µL, and column temperature 35 °C. Lambda max 240 nm was set for the detection. 0.05M, pH 4.5 sodium acetate buffer with acetonitrile in the ratio of (80:20, v/v) were used. The method was validated as per ICH guidelines and proven that the method was specific, precise, accurate, and linear. Method robustness executed by utilizing quality by design experiment and concluded method is robust. Accuracy results were found 99.6 to 100.5 for CMN, 99.8 to 100.5 for BAL, and 99.2 to 100.1 for EDTA. Precision results were obtained % RSD 1.0% for CMN, 0.8% for BAL and 0.6% for EDTA. Correlation coefficient (r2) values obtained more than 0.999 for three components. Analytical solutions are stable for 48 hrs at benchtop and refrigerator conditions. Analytical method greenness was evaluated by utilizing the GAPI, NEMI, and analytical Eco-scale tools and confirmed the method is Eco-friendly.

This publication has 24 references indexed in Scilit: