Experimental Simulation and Characterisation of Neutron Spectra for Calibrating Radiation Protection Devices

Authors:
  • MBM. Buxerolle
    M. Buxerolle
  • JCJ.L. Chartier
    J.L. Chartier
  • JKJ. Kurkdjian
    J. Kurkdjian
  • RMR. Médioni
    R. Médioni
  • MMM. Massoutié
    M. Massoutié
  • FPF. Posny
    F. Posny
  • EME. de Matos
    E. de Matos
  • MSM. Sueur
    M. Sueur

Abstract
To improve the evaluation of the dosimetric quantities, it would be advisable to calibrate the radiation protection devices in neutron spectra which are nearly like those met in practice. In order to complete the range of neutron calibrating sources, it seems useful to develop several wide spectral distributions representative of typical spectra down to thermal neutron energies. The prototype equipment of our laboratory is based on the moderation of fission neutrons using a set of different shields of different thicknesses. These broad spectral distributions have been measured by two complementary techniques; the Bonner sphere system and recoil proton spectrometry. The dosimetric quantities deduced from these spectra are compared with direct kerma measurements obtained with the 'twin-chamber' technique.