Methods of high current magnetic field generator for transcranial magnetic stimulation application
Authors:
- 21 April 2015
- journal article
- conference paper
- Published by AIP Publishing in Journal of Applied Physics
- Vol. 117 (17) , 17B319
Abstract
This paper describes the design procedures and underlying concepts of a novel High Current Magnetic Field Generator (HCMFG) with adjustable pulse width for transcranial magnetic stimulation applications. This is achieved by utilizing two different switching devices, the MOSFET and insulated gate bipolar transistor (IGBT). Results indicate that currents as high as ±1200 A can be generated with inputs of +/−20 V. Special attention to tradeoffs between field generators utilizing IGBT circuits (HCMFG1) and MOSFET circuits (HCMFG2) was considered. The theory of operation, design, experimental results, and electronic setup are presented and analyzed.Keywords
This publication has 5 references indexed in Scilit:
- Low power field generation for magneto-optic fiber-based interferometric switchesJournal of Applied Physics, 2012
- Developments in deep brain stimulation using time dependent magnetic fieldsJournal of Applied Physics, 2012
- New Designs for Deep Brain Transcranial Magnetic StimulationIEEE Transactions on Magnetics, 2011
- A Parallel 3-D impedance method to compute the Induce Current in Human Head by TMS2007
- Transcranial magnetic stimulation (TMS) in controlled treatment studies: are some “sham” forms active?Biological Psychiatry, 2000