EDXRF determination of iron during infant cereals production and its fitness for purpose
Authors:
- 20 February 2007
- journal article
- Published by Oxford University Press (OUP) in International Journal of Food Science and Technology
- Vol. 42 (5) , 551-555
Abstract
Summary: The objective of this study was to develop and validate a rapid and efficient method for the determination of iron in commercial infant cereals using energy‐dispersive X‐ray fluorescence (EDXRF) technique. Two EDXRF devices were tested: Spectro Xepos and PANalytical Minipal 2.20 samples were used for calibration of the XRF apparatus. The technique was validated using a second set of twenty other samples and comparing results of the reference method (inductively coupled plasma‐atomic emission spectroscopy, ICP‐AES) and those of the EDXRF. No significant bias was found between the two methods. Three pellets per sample are prepared under 10 tonnes of pressure and analysed. The total time of analysis for iron is less than 20 min and this technique appears to be adequate for process control during infant cereals production.Keywords
This publication has 6 references indexed in Scilit:
- Rapid quantification of iron, copper and zinc in food premixes using energy dispersive X-ray fluorescenceJournal of Food Composition and Analysis, 2005
- Wavelength‐dispersive x‐ray fluorescence measurements on organic matrices: application to milk‐based productsX-Ray Spectrometry, 2004
- ED-XRF as a Tool for Rapid Minerals Control in Milk-Based ProductsJournal of Agricultural and Food Chemistry, 2003
- Process analytical chemistry—future trends in industryAnalytical and Bioanalytical Chemistry, 2003
- Alternatives to the Median Absolute DeviationJournal of the American Statistical Association, 1993
- Alternatives to the Median Absolute DeviationJournal of the American Statistical Association, 1993