Interlaboratory benchmarking of differential scanning fluorimetry reveals sources of variability in protein thermal stability parameters
Authors:
- Sébastien BrûléSébastien Brûlé
- Maelenn ChevreuilMCMaelenn Chevreuil
- Osvaldo BurasteroOsvaldo Burastero
- Olga AbianOlga Abian
- Gregor AnderluhGregor Anderluh
- Magali Aumont-NicaiseMAMagali Aumont-Nicaise
- Martynas BagdonasMBMartynas Bagdonas
- Holly BirchenoughHBHolly Birchenough
- Alexander K. BuellAlexander K. Buell
- Audrey CrousillesACAudrey Crousilles
Preprint
Abstract
Differential scanning fluorimetry (DSF) is widely used to assess protein thermal stability and to compare protein variants, ligands, or formulations with limited sample consumption. Yet, the reproducibility of DSF-derived parameters across laboratories, instruments and analysis workflows remains poorly defined. Here, we report a two-phase interlaboratory benchmark using a shared V H H sample across 26 laboratories, 28 instruments and seven DSF instrument models. In Phase 1, participants applied their routine acquisition and analysis workflows, whereas in Phase 2, acquisition conditions were harmonised. Datasets were analysed locally and then centrally reprocessed with MoltenProt. Participant-reported apparent melting temperatures ( T m ) showed good within-instrument repeatability but moderate inter-instrument dispersion. Harmonisation of acquisition conditions did not substantially reduce this dispersion. Conversely, centralised reprocessing narrowed the distribution among retained datasets, indicating contributions from analysis workflows and data-acceptance criteria in addition to instrument-dependent factors. Comparison with differential scanning calorimetry (DSC) further showed that apparent T m depended strongly on the fluorescence readout. For this V H H, single-wavelength fluorescence at 350 nm showed the closest agreement with the DSC transition temperature, whereas the 350/330 nm ratio and barycentric mean generally yielded higher values. Furthermore, DSF-derived apparent unfolding enthalpies were more variable than apparent T m . Together, these findings show that DSF can be highly repeatable within a defined instrument and workflow, but that quantitative comparability across instruments depends on instrument-specific effects, the choice of fluorescence readout and analysis strategy. Improved comparability will require transparent reporting of acquisition and analysis conditions, explicit data-quality criteria, and well-characterised reference materials to monitor instrument-dependent offsets.Keywords
This publication has 43 references indexed in Scilit:
- Standard operating procedure for fluorescent thermal shift assay (FTSA) for determination of protein–ligand binding and protein stabilityEuropean Biophysics Journal, 2021
- A multi-laboratory benchmark study of isothermal titration calorimetry (ITC) using Ca2+ and Mg2+ binding to EDTAEuropean Biophysics Journal, 2021
- Interaction standards for biophysics: anti-lysozyme nanobodiesEuropean Biophysics Journal, 2021
- Protein formulation through automated screening of pH and buffer conditions, using the Robotein® high throughput facilityEuropean Biophysics Journal, 2021
- In‐depth interrogation of protein thermal unfolding data with MoltenProtProtein Science, 2020
- Nano differential scanning fluorimetry for comparability studies of therapeutic proteinsAnalytical Biochemistry, 2020
- Theory and applications of differential scanning fluorimetry in early-stage drug discoveryBiophysical Reviews, 2020
- The fluorescence intensities ratio is not a reliable parameter for evaluation of protein unfolding transitionsProtein Science, 2017
- Interlaboratory comparison of size measurements on nanoparticles using nanoparticle tracking analysis (NTA)Journal of Nanoparticle Research, 2013
- Extrinsic Fluorescent Dyes as Tools for Protein CharacterizationPharmaceutical Research, 2008