Electron beam quality control using an amorphous silicon EPID
Open Access
- 24 April 2009
- journal article
- Published by Wiley in Medical Physics
- Vol. 36 (5) , 1859-1866
Abstract
An amorphous siliconEPID has been investigated to determine whether it is capable of quality control constancy measurements for linear acceleratorelectron beams. The EPID grayscale response was found to be extremely linear with dose over a wide dose range and, more specifically, for exposures of 95–100 MU. Small discrepancies of up to 0.8% in linearity were found at 6 MeV (8–15 MeV showed better agreement). The shape of the beam profile was found to be significantly altered by scatter in air over the approximately 60 cm gap between the end of the applicator and the EPID. Nevertheless, relative changes in EPID-measured profile flatness and symmetry were linearly related to changes in these parameters at 95 cm focus to surface distance (FSD) measured using a 2D diode array. Similar results were obtained at 90° and 270° gantry angles. Six months of daily images were acquired and analyzed to determine whether the device is suitable as a constancy checker. EPID output measurements agreed well with daily ion chamber measurements, with a 0.8% standard deviation in the difference between the two measurement sets. When compared to weekly parallel plate chamber measurements, this figure dropped to 0.5%. A Monte Carlo(MC) model of the EPID was created and demonstrated excellent agreement between MC-calculated profiles in water and the EPID at 95 and 157 cm FSD. Good agreement was also found with measured EPID profiles, demonstrating that the EPID provides an accurate measurement of electron profiles. The EPID was thus shown to be an effective method for performing electron beam daily constancy checks.Keywords
This publication has 25 references indexed in Scilit:
- Use of an amorphous silicon EPID for measuring MLC calibration at varying gantry anglePhysics in Medicine & Biology, 2007
- Evaluation of MLC leaf positioning using a scanning liquid ionization chamber EPIDPhysics in Medicine & Biology, 2006
- Evaluation of two methods of predicting MLC leaf positions using EPID measurementsMedical Physics, 2006
- Use of an amorphous silicon electronic portal imaging device for multileaf collimator quality control and calibrationPhysics in Medicine & Biology, 2005
- Use of EPID for leaf position accuracy QA of dynamic multi‐leaf collimator (DMLC) treatmentMedical Physics, 2004
- Verification of multileaf collimator leaf positions using an electronic portal imaging deviceMedical Physics, 2002
- Assessment of flatness and symmetry of megavoltage x-ray beam with an electronic portal imaging device (EPID)Australasian Physics & Engineering Sciences in Medicine, 2002
- A test tool for the visual verification of light and radiation fields using film or an electronic portal imaging deviceMedical Physics, 1999
- Daily dosimetric quality control of the MM50 Racetrack Microtron using an electronic portal imaging deviceRadiotherapy and Oncology, 1995
- The use of an electronic portal imaging device for exit dosimetry and quality control measurementsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1995